Long-term relationship between unattended automated blood pressure and auscultatory BP measurements in hypertensive patients.

Jitka Seidlerová1,2, Jiří Ceral3, Markéta Mateřánková1, Petr König1, Ivan Řiháček4, Petra Vysočanová5, Miroslav Souček4, Jan Filipovský1,2.   

Abstract

AIMS: Unattended automated office blood pressure (uAutoOBP) has attracted more attention since SPRINT trial had been published. However, its long-term relationship to attended office blood pressure (AuscOBP) is not known.
MATERIAL AND METHODS: Stable treated hypertensive subjects were examined in four Czech academic hypertension centers. All subjects attended four clinical visits three months apart. uAutoOBP was measured with the BP Tru device; AuscOBP was measured three times with auscultatory method by the physician. 24-hour ambulatory blood pressure monitoring (ABPM) was performed within one week from the second clinical visit.
RESULTS: Data on 112 subjects aged 65.6 ± 10.8 years with mean AuscOBP 128.2 ± 12.2/78.5 ± 10.3 mm Hg are reported. Across the four clinical visits, the uAutoOBP was by 10.1/3.7 mm Hg lower than AuscOBP and the mean difference was similar during all four visits (P≥.061). Both uAutoOBP and AuscOBP had similar intra-individual variability during study follow-up as demonstrated by similar intraclass correlation coefficients (ICC, for systolic ICC = 0.50, for diastolic ICC = 0.72). However, the intra-individual variability of the systolic AuscOBP and uAutoOBP difference was high as demonstrated by low ICCs for absolute (ICC = 0.17 [95%CI, 0.09 - 0.25]) and low κ coefficients for categorized differences (κ ≤ 0.16). The main determinant of AuscOBP-uAutoOBP difference was AuscOBP level. The AuscOBP-uAutoOBP difference was poor tool to identify hypertension control categories defined on the basis of AuscOBP and ABPM.
CONCLUSIONS: Although mean AuscOBP-uAutoOBP differences were relatively similar across the four clinical visits, intra-individual variability of this difference was high. The AuscOBP-uAutoOBP difference was poor tool to identify hypertension control categories defined on the basis of AuscOBP and ABPM. Therefore, uAutoOBP cannot be used as a replacement for ABPM.

Entities:  

Keywords:  Ambulatory blood pressure monitoring; BPTru device; attended office blood pressure; blood pressure measurement; unattended automated blood pressure

Mesh:

Year:  2018        PMID: 30474412     DOI: 10.1080/08037051.2018.1540260

Source DB:  PubMed          Journal:  Blood Press        ISSN: 0803-7051            Impact factor:   2.835


  1 in total

1.  Comparative accuracies of automated and manual office blood pressure measurements in a Chinese population.

Authors:  Eric K P Lee; MengTing Zhu; Dicken C C Chan; Benjamin H K Yip; Richard McManus; Samuel Y S Wong
Journal:  Hypertens Res       Date:  2021-11-22       Impact factor: 3.872

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.