Monte Carlo investigation on the effect of air gap under bolus in post-mastectomy radiotherapy.

Eeva Boman1, Jarkko Ojala2, Maija Rossi2, Mika Kapanen2.   

Abstract

OBJECTIVES: To investigate the dosimetric effect of air gaps under bolus on skin dose for left-sided post-mastectomy radiotherapy with loco regional involvement.
METHODS: Eight patients were planned retrospectively with volume modulated arc therapy (VMAT) and conventional static Field-in-Field (FinF) methods. Three different setups were applied for the 5-mm bolus over the chest wall having 0, 5 or 10 mm air gap under the bolus. The dose calculation was performed using Monte Carlo (MC) simulation. In addition, Analytic Anisotropic Algorithm (AAA) was used to demonstrate the differences observed in clinical setting.
RESULTS: The investigated air gaps under the bolus had minimal effect on surface dose for FinF plans (relative difference ≤ 2.6%), whereas for VMAT plans the surface dose decreased 13.6% when compared to the case with no air gap. In both FinF and VMAT, the largest differences between AAA and MC were seen at the surface where AAA underestimated the dose by 1.5 Gy (p < 0.05) on average; while the dose in the target volume excluding the surface was relatively similar being on average 0.3 Gy (p > 0.05) larger with AAA than with MC calculations.
CONCLUSIONS: The surface dose was significantly lower with VMAT technique than with FinF technique. Possible air gaps under the bolus reduced the surface dose significantly further for VMAT but not for FinF treatments, which may have clinical impact on recurrence rate. AAA was shown to underestimate the surface dose when compared to MC calculation.
Copyright © 2018 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Air gaps; Breast cancer; Monte Carlo; Surface dose; VMAT

Mesh:

Year:  2018        PMID: 30471824     DOI: 10.1016/j.ejmp.2018.10.023

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  5 in total

1.  Monte Carlo Modeling of the Agility MLC for IMRT and VMAT Calculations.

Authors:  Shingo Ohira; Hideki Takegawa; Masayoshi Miyazaki; Masahiko Koizumi; Teruki Teshima
Journal:  In Vivo       Date:  2020 Sep-Oct       Impact factor: 2.155

2.  Dosimetric characteristics of a thin bolus made of variable shape tungsten rubber for photon radiotherapy.

Authors:  Katsuya Okuhata; Mikoto Tamura; Hajime Monzen; Yasumasa Nishimura
Journal:  Phys Eng Sci Med       Date:  2021-09-20

3.  3D-printed bolus ensures the precise postmastectomy chest wall radiation therapy for breast cancer.

Authors:  Xiran Wang; Jianling Zhao; Zhongzheng Xiang; Xuetao Wang; Yuanyuan Zeng; Ting Luo; Xi Yan; Zhuang Zhang; Feng Wang; Lei Liu
Journal:  Front Oncol       Date:  2022-09-02       Impact factor: 5.738

4.  Dosimetric Evaluation of Commercially Available Flat vs. Self-Produced 3D-Conformal Silicone Boluses for the Head and Neck Region.

Authors:  Stephan Pollmann; André Toussaint; Michael Flentje; Sonja Wegener; Victor Lewitzki
Journal:  Front Oncol       Date:  2022-08-10       Impact factor: 5.738

5.  Evaluation of the quality of fit of flexible bolus material created using 3D printing technology.

Authors:  Ciaran Malone; Elaine Gill; Tanith Lott; Catherine Rogerson; Sinead Keogh; Majed Mousli; Denise Carroll; Caitriona Kelly; John Gaffney; Brendan McClean
Journal:  J Appl Clin Med Phys       Date:  2022-01-20       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.