Literature DB >> 30471131

A fast, linear Boltzmann transport equation solver for computed tomography dose calculation (Acuros CTD).

Adam Wang1, Alexander Maslowski1, Todd Wareing1, Josh Star-Lack1, Taly Gilat Schmidt2.   

Abstract

PURPOSE: To improve dose reporting of CT scans, patient-specific organ doses are highly desired. However, estimating the dose distribution in a fast and accurate manner remains challenging, despite advances in Monte Carlo methods. In this work, we present an alternative method that deterministically solves the linear Boltzmann transport equation (LBTE), which governs the behavior of x-ray photon transport through an object.
METHODS: Our deterministic solver for CT dose (Acuros CTD) is based on the same approach used to estimate scatter in projection images of a CT scan (Acuros CTS). A deterministic method is used to compute photon fluence within the object, which is then converted to deposited energy by multiplying by known, material-specific conversion factors. To benchmark Acuros CTD, we used the AAPM Task Group 195 test for CT dose, which models an axial, fan beam scan (10 mm thick beam) and calculates energy deposited in each organ of an anthropomorphic phantom. We also validated our own Monte Carlo implementation of Geant4 to use as a reference to compare Acuros against for other common geometries like an axial, cone beam scan (160 mm thick beam) and a helical scan (40 mm thick beam with table motion for a pitch of 1).
RESULTS: For the fan beam scan, Acuros CTD accurately estimated organ dose, with a maximum error of 2.7% and RMSE of 1.4% when excluding organs with <0.1% of the total energy deposited. The cone beam and helical scans yielded similar levels of accuracy compared to Geant4. Increasing the number of source positions beyond 18 or decreasing the voxel size below 5 × 5 × 5 mm3 provided marginal improvement to the accuracy for the cone beam scan but came at the expense of increased run time. Across the different scan geometries, run time of Acuros CTD ranged from 8 to 23 s.
CONCLUSIONS: In this digital phantom study, a deterministic LBTE solver was capable of fast and accurate organ dose estimates.
© 2018 American Association of Physicists in Medicine.

Entities:  

Keywords:  CT dose; Monte Carlo; deterministic solver; discrete ordinates; organ dose

Mesh:

Year:  2018        PMID: 30471131      PMCID: PMC6367051          DOI: 10.1002/mp.13305

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  Technical note: Evaluation of a V-Net autosegmentation algorithm for pediatric CT scans: Performance, generalizability, and application to patient-specific CT dosimetry.

Authors:  Philip M Adamson; Vrunda Bhattbhatt; Sara Principi; Surabhi Beriwal; Linda S Strain; Michael Offe; Adam S Wang; Nghia-Jack Vo; Taly Gilat Schmidt; Petr Jordan
Journal:  Med Phys       Date:  2022-02-22       Impact factor: 4.071

2.  Validation of a deterministic linear Boltzmann transport equation solver for rapid CT dose computation using physical dose measurements in pediatric phantoms.

Authors:  Sara Principi; Yonggang Lu; Yu Liu; Adam Wang; Alex Maslowski; Todd Wareing; John Van Heteren; Taly Gilat Schmidt
Journal:  Med Phys       Date:  2021-10-29       Impact factor: 4.071

3.  Reduced Chest Computed Tomography Scan Length for Patients Positive for Coronavirus Disease 2019: Dose Reduction and Impact on Diagnostic Utility.

Authors:  Sara Principi; Stacy O'Connor; Luba Frank; Taly Gilat Schmidt
Journal:  J Comput Assist Tomogr       Date:  2022-04-08       Impact factor: 2.081

4.  Deterministic linear Boltzmann transport equation solver for patient-specific CT dose estimation: Comparison against a Monte Carlo benchmark for realistic scanner configurations and patient models.

Authors:  Sara Principi; Adam Wang; Alexander Maslowski; Todd Wareing; Petr Jordan; Taly Gilat Schmidt
Journal:  Med Phys       Date:  2020-10-20       Impact factor: 4.071

5.  XDose: toward online cross-validation of experimental and computational X-ray dose estimation.

Authors:  Philipp Roser; Annette Birkhold; Alexander Preuhs; Philipp Ochs; Elizaveta Stepina; Norbert Strobel; Markus Kowarschik; Rebecca Fahrig; Andreas Maier
Journal:  Int J Comput Assist Radiol Surg       Date:  2020-12-04       Impact factor: 2.924

  5 in total