| Literature DB >> 30467857 |
Japke F Petersen1, Liset Lansaat1, Frans Hilgers1,2, Michiel van den Brekel1,2,3.
Abstract
Entities:
Mesh:
Year: 2018 PMID: 30467857 PMCID: PMC6859417 DOI: 10.1002/lary.27645
Source DB: PubMed Journal: Laryngoscope ISSN: 0023-852X Impact factor: 3.325
Figure 1Schematic drawing and actual photo of the Provox Vega XtraSeal (Atos Medical, Hörby, Sweden) showing (A) the location of the normal esophageal flange, (B) the extra thin, angled esophageal flange, (C) and the tracheal flange.
Patient Characteristics
| No. | Percentage | |
|---|---|---|
| Gender | ||
| Male | 11 | 85% |
| Female | 2 | 15% |
| Mean age at TL (range) | 59 | 40–79 |
| Median FU in months | 117 | 7–227 |
| Indication TL | ||
| Primary | 5 | 39% |
| Salvage | 5 | 39% |
| Dysfunctional | 2 | 15% |
| Second primary | 1 | 8% |
| Origin tumor | ||
| Larynx | 10 | 77% |
| Hypopharynx | 1 | 8% |
| Other | 2 | 15% |
FU in months was calculated from date of TL to date of removal of final VP.
FU = follow‐up; TL = total laryngectomy; VP = voice prostheses.
Device Lifetime, Size, and Reason for Removal of PVX and Former VP for All Replacements
| Pt | Indication TL | Type Old VP | Size Old VP | DLT Old VP | Reason Removal Old VP | Size PVX | DLT PVX in Days | Reason Removal PVX |
|---|---|---|---|---|---|---|---|---|
| 1 | Salvage | ActiValve Light | 4,5 | 147 | PP leakage | 10 | 3 | Pt died |
| 2 | Second primary | Vega | 8 | 7 | PP leakage | 8 | 22 | TP leakage |
| 3.1 | Salvage | Vega | 6 | 31 | PP leakage | 8 | 68 | TP leakage |
| 3.2 | Salvage | PVX | 8 | NA | TP leakage | 8 | 44 | TP leakage |
| 3.3 | Salvage | Vega | 8 | NA | TP leakage | 8 | 70 | Unknown |
| 3.4 | Salvage | Vega | 8 | NA | PP leakage | 8 | 232 | Leakage NOS |
| 3.5 | Salvage | PVX | 8 | NA | Leakage NOS | 8 | 42 | TP leakage |
| 3.6 | Salvage | PVX | 8 | NA | TP leakage | 8 | 92 | TP leakage |
| 3.7 | Salvage | PVX | 6 | NA | TP leakage | 6 | 27 | TP leakage |
| 3.8 | Salvage | PVX | 8 | NA | TP leakage | 8 | 504 | NA, still in situ |
| 4.1 | Salvage | Vega + XtraFlange | 10 | 105 | PP leakage | 10 | 133 | TP leakage |
| 4.2 | Salvage | PVX | 10 | NA | TP leakage | 10 | 223 | TP leakage |
| 4.3 | Salvage | PVX | 10 | NA | TP leakage | 10 | 835 | NA, still in situ |
| 5 | DF larynx | Vega | 8 | 99 | PP leakage | 8 | 34 | Leakage NOS |
| 6.1 | Primary | Vega | 10 | 28 | PP leakage | 10 | 63 | Leakage NOS |
| 6.2 | Primary | PVX | 10 | NA | Leakage NOS | 6 | 41 | Leakage NOS |
| 7.1 | Salvage | Vega + XtraFlange | 6 | 25 | PP leakage | 6 | 79 | TP leakage |
| 7.2 | Salvage | PVX | 8 | NA | Unknown | 8 | 62 | TP leakage |
| 7.3 | Salvage | PVX | 8 | NA | Unknown | 8 | 39 | Pt died |
| 8 | Salvage | Vega | 6 | 246 | PP leakage | 6 | 91 | TP leakage |
| 9 | Primary | Vega | 8 | 4 | VP lost | 8 | 15 | PP leakage |
| 10 | Primary | Vega | 8 | 38 | PP leakage | 8 | 156 | TP leakage |
| 11.1 | Primary | Vega | 8 | 20 | Wide TEP tract | 8 | 4 | VP lost |
| 11.2 | Primary | PVX | 8 | NA | VP lost | 8 | 28 | Surgical revision |
| 12 | Dysfunctional larynx | ActiValve Light | 8 | 60 | PP leakage | 8 | 36 | TP leakage |
| 13 | Primary | Vega | 8 | 222 | PP leakage | 8 | 232 | Voice problems |
All VP used in this study are Atos Medical Products (Hörby, Sweden).
DF larynx = dysfunctional larynx; DLT = device lifetime; NOS = not otherwise specified; old VP = voice prostheses replaced with Provox Vega XtraSeal; PP leakage = periprosthetic leakage; Pt = patient; PVX = Provox Vega XtraSeal; TL = total laryngectomy; TP leakage = transprosthetic leakage; VP = voice prostheses.