Literature DB >> 30466818

Timely recognition of total elbow and radial head arthroplasty adverse events: an analysis of reports to the US Food and Drug Administration.

Jeremy S Somerson1, Frederick A Matsen2.   

Abstract

BACKGROUND: Recent recalls of several commonly used elbow arthroplasty implants have prompted interest in the modes by which elbow implants fail and the timing of reports of these failures.
METHODS: We reviewed the adverse event reports to the US Food and Drug Administration's Manufacturer and User Facility Device Experience (MAUDE) database from 2012 to 2015 regarding elbow arthroplasty to determine the event date and the type of each adverse event.
RESULTS: Among 179 total elbow adverse event reports, the most common modes of failure were implant dissociation (23%), loosening (22%), and infection (16%). The most common modes of failure among 58 radial head replacement reports were component dissociation (19%) and linkage screw failure (19%). The percentage distribution of adverse event types differed among different arthroplasty systems and from that reported in published reviews of elbow arthroplasty. Three implant recalls were implemented 2, 5, and 9 years after the first adverse event report in the MAUDE database. For 2 of the recalls, the first reports of the device failures were published 2 and 5 years after the first MAUDE reports.
CONCLUSIONS: The MAUDE database is a publicly funded and publicly available means by which surgeons can identify adverse events for the prostheses they use before such information becomes available through journal publication or recall notification. In this study, MAUDE data revealed a higher relative frequency of mechanical dissociation of elbow implants than what has been represented in the literature. Early identification of these adverse events may help surgeons by informing their implant selection and surgical technique.
Copyright © 2018 Journal of Shoulder and Elbow Surgery Board of Trustees. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Elbow arthroplasty; MAUDE database; dissociation; failure; loosening; recall

Mesh:

Year:  2018        PMID: 30466818     DOI: 10.1016/j.jse.2018.08.043

Source DB:  PubMed          Journal:  J Shoulder Elbow Surg        ISSN: 1058-2746            Impact factor:   3.019


  2 in total

1.  CORR Insights®: What Factors Are Associated with Reoperation after Operative Treatment of Terrible Triad Injuries?

Authors:  W Angus Wallace
Journal:  Clin Orthop Relat Res       Date:  2021-01-01       Impact factor: 4.755

2.  Adverse events related to total ankle replacement devices: an analysis of reports to the United States Food and Drug Administration.

Authors:  Karim Mahmoud; Sreenivasulu Metikala; Kathryn M O'Connor; Daniel C Farber
Journal:  Int Orthop       Date:  2021-02-11       Impact factor: 3.075

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.