Accuracy of an automated blood-gas analyser operated by untrained staff.

B D Minty, A M Barrett.   

Abstract

Performance of the IL 613 automatic blood-gas analyser has been assessed using a group of 100 "analysts" with no previous training or experience in the use of the instrument. Test material consisted of blood equilibrated to a known PO2 and PCO2 in a tonometer: pH estimations were carried out on thawed aliquots of a large batch of frozen serum which were then equilibrated to a known PCO2. Eighty-six per cent of analyses were within acceptable limits of error. The largest proportion of errors was in the measurement of pH. Satisfactory results were obtained in 98% of the analyses of PO2 and PCO2. Eighty-eight per cent of operators were able to use the analyser after instruction lasting less than 1 min. These results were significantly better than those obtained in a regional survey of 16 blood-gas laboratories, staffed by trained technicians.

Entities:  

Mesh:

Substances:

Year:  1978        PMID: 30464     DOI: 10.1093/bja/50.10.1031

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  3 in total

1.  Electrolyte measurement by clinicians.

Authors:  S M Willatts; K Myerson
Journal:  Intensive Care Med       Date:  1987       Impact factor: 17.440

2.  Clinical biochemistry nearer the patient.

Authors:  V Marks
Journal:  Br Med J (Clin Res Ed)       Date:  1983-04-09

3.  Oxygen tension measurement using an automatic blood gas analyser.

Authors:  J Becket; C Orchard; M K Chakrabarti; G M Hall; I D Gillies; P J Bourdillon
Journal:  J R Soc Med       Date:  1981-08       Impact factor: 18.000

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.