Fiberoptic versus Digital: A Comparison of Durability and Cost Effectiveness of the Two Flexible Ureteroscopes.

Mustafa Zafer Temiz1, Aykut Colakerol2, Kasim Ertas2, Murat Tuken2, Emrah Yuruk2.   

Abstract

AIMS: We aimed to evaluate the durability and cost effectiveness of the latest digital flexible ureterescope by comparing it with the conventional fiberoptic one.
MATERIALS AND METHODS: Data of patients who underwent retrograde intrarenal surgery between January 2013 and December 2014 were collected. Fiberoptic Flex-X2 or digital Cobra vision flexible ureteroscopes were used for the procedures. The comparison of both ureteroscopes was performed in terms of patient and stone characteristics, operative outcomes, durability, and cost effectiveness.
RESULTS: A total of 105 patients were evaluated for the study. The patient and stone characteristics and operative outcomes were similar between the groups. Overall, 54 and 51 procedures were performed using Flex-X2 and Cobra vision, respectively, before they were sent for renovation. The purchase prices were USD 29,500 for Flex-X2 and USD 58,000 for Cobra vision. Costs of per case were determined as USD 549.29 for Flex-X2 and as USD 1,137.25 for Cobra vision. Per minute working time costs were USD 772.04 and 1,471.33 for Flex-X2 and Cobra vision respectively.
CONCLUSIONS: The digital Cobra vision has high costs without any difference in durability as compared to Flex-X2. Moreover, it has no benefit over Flex-X2 in terms of surgical outcomes.
© 2018 S. Karger AG, Basel.

Entities:  

Keywords:  Cost-benefit analysis; Ureteroscopes; Ureteroscopy

Mesh:

Year:  2018        PMID: 30463076     DOI: 10.1159/000494385

Source DB:  PubMed          Journal:  Urol Int        ISSN: 0042-1138            Impact factor:   2.089


  3 in total

Review 1.  Cost comparison of single-use versus reusable flexible ureteroscope: A systematic review.

Authors:  Eugenio Ventimiglia; Alvaro Jiménez Godínez; Olivier Traxer; Bhaskar K Somani
Journal:  Turk J Urol       Date:  2020-08-25

Review 2.  Handling and protecting your flexible ureteroscope: how to maximise scope usage.

Authors:  Khaled Hosny; Jennifer Clark; Shalom J Srirangam
Journal:  Transl Androl Urol       Date:  2019-09

3.  Stone removing efficiency and safety comparison between single use ureteroscope and reusable ureteroscope: a systematic review and meta-analysis.

Authors:  Yu-Cheng Ma; Zhong-Yu Jian; Xi Jin; Hong Li; Kun-Jie Wang
Journal:  Transl Androl Urol       Date:  2021-04
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.