Evaluating the impact of extended field-of-view CT reconstructions on CT values and dosimetric accuracy for radiation therapy.

Joey P Cheung1, Erin Shugard1, Nilesh Mistry2, Jean Pouliot1, Josephine Chen1.   

Abstract

PURPOSE: Wide bore CT scanners use extended field-of-view (eFOV) reconstruction algorithms to attempt to recreate tissue truncated due to large patient habitus. Radiation therapy planning systems rely on accurate CT numbers in order to correctly plan and calculate radiation dose. This study looks at the impact of eFOV reconstructions on CT numbers and radiation dose calculations in real patient geometries.
METHODS: A large modular phantom based on real patient geometries was created to surround a CIRS Model 062M phantom. The modular sections included a smooth patient surface, a skin fold in the patient surface, and the addition of arms for simulation of the patient in arms up or arms down position. This phantom was used to evaluate the accuracy of CT numbers for three extended FOV algorithms implemented on Siemens CT scanners: eFOV, HDFOV, and HDProFOV. Six different configurations of the phantoms were scanned and images were reconstructed for the three different extended FOV algorithms. The CIRS phantom inserts and overall phantom geometry were contoured in each image, and the Hounsfield units (HU) numbers were compared to an image of the phantom within the standard scan FOV (sFOV) without the modular sections. To evaluate the effect on dose calculations, six radiotherapy patients previously treated at our institution (three head and neck and three chest/pelvis) whose body circumferences extended past the 50 cm sFOV in the treatment planning CT were used. Images acquired on a Siemens Sensation Open scanner were reconstructed using sFOV, eFOV and HDFOV algorithms. A physician and dosimetrist identified the radiation target, critical organs, and external patient contour. A benchmark CT was created for each patient, consisting of an average of the 3 CT reconstructions with a density override applied to regions containing truncation artifacts. The benchmark CT was used to create an optimal radiation treatment plan. The plan was copied onto each CT reconstruction without density override and dose was recalculated.
RESULTS: Tissue extending past the sFOV impacts the HU numbers for tissues inside and outside the sFOV when using extended FOV reconstructions. On average, the HU for all CIRS density inserts in the arms up (arms down) position varied by 43 HU (67 HU), 39 HU (73 HU), and 18 HU (51 HU) for the eFOV, HDFOV, and HDProFOV scans, respectively. In the patient dose calculations, patients with a smooth patient contour had the least deviation from the benchmark in the HDFOV (0.1-0.5%) compared to eFOV (0.4-1.8%) reconstructions. In cases with large amounts of tissue and irregular skin folds, the eFOV deviated the least from the benchmark (range 0.2-0.6% dose difference) compared to HDFOV (range 1.3-1.8% dose difference).
CONCLUSIONS: All reconstruction algorithms demonstrated good CT number accuracy in the center of the image. Larger artifacts are seen near and extending outside the scan FOV, however, dose calculations performed using typical beam arrangements using the extended FOV reconstructions were still mostly within 2.5% of best estimated reference values.
© 2018 American Association of Physicists in Medicine.

Entities:  

Keywords:  CT simulation; extended field of view; radiation therapy; treatment planning

Mesh:

Year:  2018        PMID: 30457170     DOI: 10.1002/mp.13299

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Soft tissue variations influence HR-pQCT density measurements in a spatially dependent manner.

Authors:  Po-Hung Wu; Tanvi Gupta; Hanling Chang; Dimitry Petrenko; Anne Schafer; Galateia Kazakia
Journal:  Bone       Date:  2020-06-27       Impact factor: 4.398

2.  Evaluation of the high definition field of view option of a large-bore computed tomography scanner for radiation therapy simulation.

Authors:  Richard Y Wu; Tyler D Williamson; Narayan Sahoo; Trang Nguyen; Shane M Ikner; Amy Y Liu; Paul G Wisdom; MingFu Lii; Rachel A Hunter; Paola E Alvarez; G Brandon Gunn; Steven J Frank; Yoshifumi Hojo; X Ronald Zhu; Michael T Gillin
Journal:  Phys Imaging Radiat Oncol       Date:  2020-03-26

3.  Impact of computed tomography (CT) reconstruction kernels on radiotherapy dose calculation.

Authors:  Irina Vergalasova; Michael McKenna; Ning Jeff Yue; Meral Reyhan
Journal:  J Appl Clin Med Phys       Date:  2020-09-05       Impact factor: 2.102

4.  Effects of image distortion and Hounsfield unit variations on radiation treatment plans: An extended field-of-view reconstruction in a large bore CT scanner.

Authors:  Yong-Ki Bae; Jeong-Woo Lee; Semie Hong
Journal:  Sci Rep       Date:  2020-01-16       Impact factor: 4.379

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.