| Literature DB >> 30455866 |
Yussr M Ibrahim1, Nicole R Marques1, Carlos R Garcia1, Michael Salter1, Christopher McQuitty1, Michael Kinsky1, Mindy Juan2, Achiau Ludomirsky3.
Abstract
BACKGROUND: Intravascular air embolism (AE) is a preventable but potentially catastrophic complication caused by intravenous tubing, trauma, and diagnostic and surgical procedures. The potentially fatal risks of arterial AE are well-known, and emerging evidence demonstrates impact of venous AEs on inflammatory response and coagulation factors. A novel FDA-approved in-line air detection and purging system was used to detect and remove air caused by administering a rapid fluid bolus during surgery.Entities:
Keywords: Air embolism; Anesthesia; ClearLine IV; Fluid warmers; Intravenous
Year: 2018 PMID: 30455866 PMCID: PMC6223009 DOI: 10.1186/s13741-018-0104-9
Source DB: PubMed Journal: Perioper Med (Lond) ISSN: 2047-0525
Fig. 1Setup for IV fluid administration. Fluid administration setup with ClearLine IV. Ultrasound sensing technology detects and diverts air from the IV line to a collection bag through opening of a valve to a waste reservoir. The fluid is redirected to the patient line after the air is entrapped through another valve opening or can be bypassed (blue lines). IV: intravenous
Fig. 2Air quantification in the right atrium by optical densitometry (OD). ClearLine IV (left) versus control (right). a–c TEE images during bolus. LA: left atria, RA: right atria, RVOT: right ventricular outflow tract
Baseline demographics
| Gender | Age | Race/ethnicity | |
|---|---|---|---|
| Subject 1 | Male | 72 | White/non-Hispanic |
| Subject 2 | Male | 61 | White/non-Hispanic |
| Subject 3 | Male | 60 | Black/non-Hispanic |
| Subject 4 | Male | 65 | White/non-Hispanic |
| Subject 5 | Male | 49 | White/Hispanic |
| Subject 6 | Female | 73 | White/non-Hispanic |
Air quantification by optical densitometry (OD)
| ClearLine IV mean ± SD | Control mean ± SD | ||
|---|---|---|---|
| Fluid flow rates (mL/min) | 145 ± 22 | 137 ± 45 | NS |
| Mean RA air by OD (%) | 16 ± 3 | 46 ± 7 | |
| Air clearance time (s) | 3.7 ± 1.2 | 5.3 ± 1.3 |
Fig. 3Air quantification as measured in the right atrium (RA) by TEE images and optical densitometry (OD). Less air was quantified in the RA following boluses administered in-line with the ClearLine IV system. TEE: transesophageal echocardiography