PATIENT DOSE IN DIGITAL RADIOGRAPHY UTILISING BMI CLASSIFICATION.

Vasileios I Metaxas1, Gerasimos A Messaris1, Aristea N Lekatou1, Theodore G Petsas2, George S Panayiotakis1.   

Abstract

Dose audit is important towards optimisation of patients' radiation protection in diagnostic radiography. In this study, the effect of the body mass index (BMI) on radiation dose received by 1869 adult patients undergoing chest, abdomen, lumbar spine, kidneys and urinary bladder (KUB) and pelvis radiography in an X-ray room with a digital radiography system was investigated. Patients were categorised into three groups (normal, overweight and obese) based on the BMI values. The patients' entrance surface air kerma (ESAK) and the effective dose (ED) were calculated based on the X-ray tube output, exposure parameters and technical data, as well as utilising appropriate conversion coefficients of the recorded kerma area product (KAP) values. The local diagnostic reference levels (LDRLs) were established at the 75th percentile of the distribution of ESAK and KAP values. Statistically, a significant increase was found in ESAK, KAP and ED values, for all examinations, both for overweight and obese patients compared to normal patients (Mann-Whitney test, p < 0.0001). Regarding the gender of the patients, a statistically significant increase was found in the dose values for male patients compared to female patients, except for the chest LAT examinations (Mann-Whitney test, p = 0.06). The percentage increase for chest PA, chest LAT, abdomen AP, lumbar spine AP, lumbar spine LAT, pelvis AP and KUB AP in overweight patients was 75%, 100%, 136%, 130%, 70%, 66% and 174% for median ESAK, 67%, 81%, 135%, 134%, 85%, 63% and 172% for median KAP, as well as 89%, 54%, 146%, 138%, 82%, 57% and 183% for median ED values, respectively. For obese patients, the corresponding increases were 200%, 186%, 459%, 345%, 203%, 150% and 785% for median ESAK, 200%, 185%, 423%, 357%, 227%, 142% and 597% for median KAP, as well as 222%, 156%, 446%, 363%, 218%, 136% and 625% for median ED. The corresponding LDRLs for overweight patients were 0.17 mGy, 1.21 mGy, 3.74 mGy, 7.70 mGy, 7.99 mGy, 4.07mGy, 5.03 mGy and 0.13 Gy cm2, 0.69 Gy cm2, 2.35 Gy cm2, 2.10 Gy cm2, 2.59 Gy cm2, 2.13 Gy cm2, 2.49 Gy cm2 in terms of ESAK and KAP values, respectively, while in the case of obese patients were 0.28 mGy, 1.82 mGy, 7.26 mGy, 15.10 mGy, 13.86 mGy, 6.89 mGy, 13.40 mGy and 0.21 Gy cm2, 1.10 Gy cm2, 4.68 Gy cm2, 4.01 Gy cm2, 4.80 Gy cm2, 3.27 Gy cm2, 6.02 Gy cm2, respectively. It can be concluded that overweight and obese patients received a significantly increased radiation dose. Careful adjustment of imaging protocols is needed for these patients to reduce patient dose, while keeping the image quality at an acceptable level. Additional studies need to be conducted for these patient groups, that could further contribute to the development of radiation protection culture in diagnostic radiography.
© The Author(s) 2018. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Entities:  

Mesh:

Year:  2019        PMID: 30452729     DOI: 10.1093/rpd/ncy194

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  4 in total

1.  Comparative outcomes of Inferior Vena Cava filters placed at bedside using digital radiography versus conventional fluoroscopy.

Authors:  John A Walker; Matthew Milam; Jorge E Lopera
Journal:  J Interv Med       Date:  2021-05-12

2.  Clinical utility of postprocessed low-dose radiographs in skeletal imaging.

Authors:  Johannes Kolck; Katharina Ziegeler; Thula Walter-Rittel; Kay Geert A Hermann; Bernd Hamm; Alexander Beck
Journal:  Br J Radiol       Date:  2022-01-05       Impact factor: 3.039

3.  Radiation exposure in endovascular stroke treatment of acute basilar artery occlusions-a matched-pair analysis.

Authors:  Charlotte S Weyland; Arne Potreck; Ulf Neuberger; Markus A Möhlenbruch; Simon Nagel; Peter A Ringleb; Martin Bendszus; Johannes A R Pfaff
Journal:  Neuroradiology       Date:  2020-07-10       Impact factor: 2.804

4.  Influence of the use of various imaging units and projections on the radiation dose received by children during chest digital radiography.

Authors:  Hongrong Xu; Kaiping Huang; Bo Liu; Jinhua Cai; Huan Zheng; Helin Zheng; Qiurui Yang; Changhong Yao
Journal:  PLoS One       Date:  2021-08-05       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.