Literature DB >> 30450312

Preventing condensation of objective lens in noncontact wide-angle viewing systems during vitrectomy.

Jung Pil Lee1, Jinsoo Kim2, Inwon Park2, Ho Ra3, Soonil Kwon2.   

Abstract

AIM: To assess the optimal conditions for preventing condensation of objective lens during vitrectomy with noncontact wide-angle viewing systems (WAVSs).
METHODS: We explored the effectiveness of the coating with ophthalmic viscoelastic device (OVDs) on the corneal surface and the soaking the objective lens in warm-saline for preventing condensation of objective lens. First, to find the optimal soaking time to keep the objective lens warm, we measured the temperature of objective lens every minute after soaking in warm saline. Second, to find optimal distance between cornea and objective lens, which provide as wide a view as possible and less condensation at the same time, we measured the condensation time with different distances. With the obtained optimal soaking time and distance, we explored the effect of coating cornea with OVDs and soaking objective lens in warm saline on condensation time.
RESULTS: One and 5min of soaking in warm saline was most effective for keeping the lens warm enough (45.1°C±2.1°C for 1min and 46.4°C ±1.0°C for 5min, P=0.109). The mean condensation times for the control group at 1, 3, and 5 mm from corneal surface to objective lens were 1±0.4, 4±1.4, 190±26.1s, respectively, thus 5 mm was most optimal distance for vitrectomy with WAVSs. For the OVD coating group, the mean condensation times were 1.5±0.3, 13±1.4, and 200±23.9s at 1, 3, and 5 mm distance and borderline significant compared with control group (P=0.068, 0.051, and 0.063, respectively). With the 1-minute warm saline soaking group, the mean condensation time were extended to 188±34.4, 416±65.7, and 600±121.3s at 1, 3, and 5 mm distance and statistically significant compared with control (P=0.043, 0.041 and 0.043, respectively).
CONCLUSION: OVD coating on corneal surface shows no difference on condensation time with control group. However, soaking the objective lens in warm saline revealed statistically significant extension of condensation time compared to control group. Therefore, keeping the objective lens warm with soaking in warm saline is a simple but effective to prevent condensation of objective lens during vitrectomy. The thermodynamics between objective lens and cornea during vitrectomy warrants further investigation.

Entities:  

Keywords:  condensation; vitrectomy; warm saline; wide-angle viewing systems

Year:  2018        PMID: 30450312      PMCID: PMC6232337          DOI: 10.18240/ijo.2018.11.12

Source DB:  PubMed          Journal:  Int J Ophthalmol        ISSN: 2222-3959            Impact factor:   1.779


  19 in total

1.  A new, non-contact wide field viewing system for vitreous surgery.

Authors:  Maurice B Landers; Gholam A Peyman; Izak F Wessels; Paul Whalen; Virgilio Morales
Journal:  Am J Ophthalmol       Date:  2003-07       Impact factor: 5.258

2.  Combining a contact lens and wide-angle viewing system for a wider fundus view.

Authors:  Masahito Ohji; Eijiroh Tada; Hideyuki Futamura
Journal:  Retina       Date:  2011-10       Impact factor: 4.256

3.  ANTIFOGGING DEVICE TO PREVENT MOISTURE CONDENSATION DURING VITRECTOMY WITH NONCONTACT WIDE-FIELD VIEWING SYSTEM.

Authors:  Shunji Kusaka; Kuniko Tachibana; Daishi Tsujioka; Fumika Hotta; Hiroshi Eguchi; Yoshikazu Shimomura
Journal:  Retina       Date:  2017-06       Impact factor: 4.256

Review 4.  Wide-angle viewing system.

Authors:  Makoto Inoue
Journal:  Dev Ophthalmol       Date:  2014-08-26

5.  A binocular indirect ophthalmomicroscope (BIOM) for non-contact wide-angle vitreous surgery.

Authors:  M Spitznas
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  1987       Impact factor: 3.117

6.  A new device for noncontact wide-angle viewing of the fundus during vitrectomy.

Authors:  E H Bovey; M Gonvers
Journal:  Arch Ophthalmol       Date:  1995-12

7.  Thermodynamics of vitreoretinal surgery.

Authors:  Mario R Romano; Jose L Vallejo-Garcia; Vito Romano; Martina Angi; Paolo Vinciguerra; Ciro Costagliola
Journal:  Curr Eye Res       Date:  2012-12-11       Impact factor: 2.424

Review 8.  Vitreoretinal instruments: vitrectomy cutters, endoillumination and wide-angle viewing systems.

Authors:  Paulo Ricardo Chaves de Oliveira; Alan Richard Berger; David Robert Chow
Journal:  Int J Retina Vitreous       Date:  2016-12-05

9.  New type of antidrying lens for vitreous surgery with a noncontact wide-angle viewing system.

Authors:  Tomoyuki Chihara; Mihori Kita
Journal:  Clin Ophthalmol       Date:  2013-02-15

10.  Comparison of scleral buckling and vitrectomy using wide angle viewing system for rhegmatogenous retinal detachment in patients older than 35 years.

Authors:  Sung Who Park; Han Jo Kwon; Ho Yun Kim; Ik Soo Byon; Ji Eun Lee; Boo Sup Oum
Journal:  BMC Ophthalmol       Date:  2015-09-11       Impact factor: 2.209

View more
  2 in total

1.  Use of a hand-held fan to prevent fogging of the wide angle viewing system during vitrectomy.

Authors:  Murat Kucukevcilioglu; Ali Hakan Durukan
Journal:  Int Ophthalmol       Date:  2021-04-27       Impact factor: 2.031

2.  Wide-angle viewing lens fogging: solving a common problem in vitreoretinal surgery.

Authors:  Giuseppe Demarinis; Filippo Tatti; Enrico Peiretti
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2020-11-18       Impact factor: 3.117

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.