| Literature DB >> 30450021 |
Frederic Lange1, Luke Dunne1, Lucy Hale2,3, Ilias Tachtsidis1.
Abstract
We present a multiwavelength, multichannel, time-domain near-infrared spectroscopy system named MAESTROS. This instrument can measure absorption and scattering coefficients and can quantify the concentrations of oxy- and deoxy-haemoglobin ([HbO2], [HHb]), and oxidation state of cytochrome-c-oxidase ([oxCCO]). This system is composed of a supercontinuum laser source coupled with two acousto-optic tuneable filters. The light is collected by four photomultipliers tubes, connected to a router to redirect the signal to a single time-correlated single-photon counting card. The interface between the system and the tissue is based on optical fibres. This arrangement allows us to resolve up to 16 wavelengths, within the range of 650-900 nm, at a sampling rate compatible with the physiology (from 0.5 to 2 Hz). In this paper, we describe the system and assess its performance based on two specifically designed protocols for photon migration instruments, the basic instrument protocol and nEUROPt protocols, and on a well characterized liquid phantom based on Intralipid and water. Then, the ability to resolve [HbO2 ], [HHb], and [oxCCO] is demonstrated on a homogeneous liquid phantom, based on blood for [HbO2], [HHb], and yeast for [oxCCO]. In the future, the system could be used to monitor brain tissue physiology.Entities:
Keywords: Time domain measurements; biomedical engineering; laser biomedical applications; photomultipliers; scattering; spectroscopy
Year: 2018 PMID: 30450021 PMCID: PMC6054019 DOI: 10.1109/JSTQE.2018.2833205
Source DB: PubMed Journal: IEEE J Sel Top Quantum Electron ISSN: 1077-260X Impact factor: 4.544