| Literature DB >> 30441028 |
Aref Smiley, Steve J A Majerus, Ian S McAdams, Brett Hanzlicek, Dennis Bourbeau, Margot S Damaser.
Abstract
New research and diagnosis tools are needed to continuously measure bowel state and activity. We investigated functionality of several sensors in vivo and in vitro. Five sensor types, including pressure, infrared, color, conductivity and capacitance, were tested to validate functionality inside the colon. Initial wired prototypes were tested and calibrated in benchtop testing and then inserted intraluminally into pig colon and rectum in three acute surgical procedures. The results from both benchtop and in-vivo testing correlate and indicate that pressure, conductivity, and capacitance measurements could provide information on the state of the bowel and its activity.Entities:
Mesh:
Year: 2018 PMID: 30441028 PMCID: PMC6889208 DOI: 10.1109/EMBC.2018.8512948
Source DB: PubMed Journal: Annu Int Conf IEEE Eng Med Biol Soc ISSN: 2375-7477