Literature DB >> 30425114

Randomised study comparing heart rate measurement in newly born infants using a monitor incorporating electrocardiogram and pulse oximeter versus pulse oximeter alone.

Madeleine C Murphy1,2,3, Laura De Angelis4, Lisa K McCarthy1,5, Colm Patrick Finbarr O'Donnell1,2,3,5.   

Abstract

AIM: To determine whether IntelliVue (ECG plus Masimo pulse oximeter (PO)) measures heart rate (HR) in low-risk newborns more quickly than Nellcor PO (PO alone).
METHODS: Unmasked parallel group randomised (1:1) study.
RESULTS: We studied 100 infants, 47 randomised to IntelliVue, 53 to Nellcor. Time to first HR was shorter with IntelliVue ECG than Nellcor (median (IQR) 24 (19, 39) vs 48 (36, 69) s, p<0.001). There was no difference in time to display both HR and SpO2 (52 (47, 76) vs 48 (36, 69) s, p=0.507). IntelliVue PO displayed initial bradycardia more often than the Nellcor (55% vs 6%). Infants monitored with IntelliVue were handled more frequently and for longer.
CONCLUSIONS: IntelliVue ECG displayed HR more quickly than Nellcor PO. IntelliVue PO often displayed initial bradycardia. Infants monitored with IntelliVue were handled more often. Study of ECG in high-risk infants is warranted. © Author(s) (or their employer(s)) 2019. No commercial re-use. See rights and permissions. Published by BMJ.

Entities:  

Keywords:  electrocardiogram; heart rate; infant; newborn; pulse oximeter

Mesh:

Year:  2018        PMID: 30425114     DOI: 10.1136/archdischild-2017-314366

Source DB:  PubMed          Journal:  Arch Dis Child Fetal Neonatal Ed        ISSN: 1359-2998            Impact factor:   5.747


  3 in total

Review 1.  Oxygen saturation and heart rate in healthy term and late preterm infants with delayed cord clamping.

Authors:  Inmaculada Lara-Cantón; Shiraz Badurdeen; Janneke Dekker; Peter Davis; Calum Roberts; Arjan Te Pas; Máximo Vento
Journal:  Pediatr Res       Date:  2022-01-07       Impact factor: 3.756

2.  Physiologically based cord clamping for infants ≥32+0 weeks gestation: A randomised clinical trial and reference percentiles for heart rate and oxygen saturation for infants ≥35+0 weeks gestation.

Authors:  Shiraz Badurdeen; Peter G Davis; Stuart B Hooper; Susan Donath; Georgia A Santomartino; Alissa Heng; Diana Zannino; Monsurul Hoq; C Omar F Kamlin; Stefan C Kane; Anthony Woodward; Calum T Roberts; Graeme R Polglase; Douglas A Blank
Journal:  PLoS Med       Date:  2022-06-23       Impact factor: 11.613

3.  Heart rate detection properties of dry-electrode ECG compared to conventional 3-lead gel-electrode ECG in newborns.

Authors:  Hanne Pike; Joar Eilevstjønn; Peder Bjorland; Jørgen Linde; Hege Ersdal; Siren Rettedal
Journal:  BMC Res Notes       Date:  2021-05-01