Modified method for prophylaxis against unishunt system complications with presentation of total intraventricular migration of unisystem ventriculoperitoneal shunt.

H T Abou el Nasr1.   

Abstract

This report is a case of shunted hydrocephalus in which the Unishunt system was placed without the clips (lock and slip). Upward migration of the whole system into the lateral ventricle was discovered 1 month after surgery. The shunt was reinserted, placing the proximal "lock-clip" under the cranial flap and the distal "slip-clip" subcutaneously at the abdominal area. The skin over the slip-clip ulcerated, so it was repositioned and fixed deep in the abdominal muscle, extraperitoneally. Different factors caused the migration: negative sucking intraventricular pressure; pushing, positive intra-abdominal pressure; the course of the subcutaneous tract of the tube not being straight; incorrect fixation of the ends of the system; faulty handling of the baby. Choosing the type of shunt, its pressure opening, antisiphon device, uni- or multi-component, and following a straight line between the ventricular and abdominal inlets are all important factors that guard against migration and assure fixation. The modified fixation method using clips is recommended as a routine technique, especially in emaciated babies, to guard against skin ulceration.

Entities:  

Mesh:

Year:  1988        PMID: 3042135     DOI: 10.1007/BF00271393

Source DB:  PubMed          Journal:  Childs Nerv Syst        ISSN: 0256-7040            Impact factor:   1.475


  2 in total

1.  Peritoneal bypass in the treatment of hydrocephalus: historical review and abdominal complications.

Authors:  R I Davidson
Journal:  J Neurol Neurosurg Psychiatry       Date:  1976-07       Impact factor: 10.154

2.  Complications of ventriculo-peritoneal shunting and a critical comparison of the three-piece and one-piece systems.

Authors:  A J Raimondi; J S Robinson; K Kuwawura
Journal:  Childs Brain       Date:  1977
  2 in total
  11 in total

1.  Upward migration of a peritoneal catheter following ventriculoperitoneal shunt.

Authors:  Kyung Rae Cho; Je Young Yeon; Hyung Jin Shin
Journal:  J Korean Neurosurg Soc       Date:  2013-06-30

2.  Total intraventricular migration of unisystem ventriculo-peritoneal shunt.

Authors:  M S Eljamel; S Sharif; C N Pidgeon
Journal:  Acta Neurochir (Wien)       Date:  1995       Impact factor: 2.216

3.  Intra-ventricular migration of a subcutaneous reservoir after neuroendoscopy.

Authors:  Daniela Falk; Marcus O Pinsker; H Maximilian Mehdorn; Arya Nabavi
Journal:  Childs Nerv Syst       Date:  2017-10-03       Impact factor: 1.475

Review 4.  Shunts vs endoscopic third ventriculostomy in infants: are there different types and/or rates of complications? A review.

Authors:  C Di Rocco; L Massimi; G Tamburrini
Journal:  Childs Nerv Syst       Date:  2006-10-20       Impact factor: 1.475

5.  Proximal migration and subcutaneous coiling of a peritoneal catheter: report of two cases.

Authors:  K J Kim; K C Wang; B K Cho
Journal:  Childs Nerv Syst       Date:  1995-07       Impact factor: 1.475

6.  Distal ventriculoperitoneal shunt catheter tightly coiled around the valve in the absence of a subgaleal cerebrospinal fluid collection: illustrative case.

Authors:  Goichiro Tamura; Kerry A Vaughan; Sara Breitbart; Helen M Branson; George M Ibrahim
Journal:  J Neurosurg Case Lessons       Date:  2021-05-17

7.  Intraventricular migration of a Rickham reservoir: endoscopic retrieval.

Authors:  Kanna K Gnanalingham; Jesus Lafuente; William Harkness
Journal:  Childs Nerv Syst       Date:  2003-05-10       Impact factor: 1.475

8.  Total intracranial shunt migration.

Authors:  Vikas Naik; Manoj Phalak; Poodipedi Sarat Chandra
Journal:  J Neurosci Rural Pract       Date:  2013-01

9.  Shunt malfunction due to proximal migration and subcutaneous coiling of a peritoneal catheter.

Authors:  Amit Agarwal; Anand Kakani
Journal:  J Neurosci Rural Pract       Date:  2010-07

10.  Retrograde migration and subcutaneous coiling of the peritoneal catheter of a ventriculoperitoneal shunt in a cat.

Authors:  Sinead E Bennett; Sebastien Behr
Journal:  JFMS Open Rep       Date:  2016-05-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.