Literature DB >> 30421021

Minimizing individual variations in arterial enhancement on coronary CT angiographs using "contrast enhancement optimizer": a prospective randomized single-center study.

Yoriaki Matsumoto1,2, Toru Higaki3, Takanori Masuda4,3, Tomoyasu Sato5, Yuko Nakamura3, Fuminari Tatsugami3, Kazuo Awai3.   

Abstract

OBJECTIVES: To investigate the clinical utility of our newly developed contrast enhancement optimizer (CEO) software for coronary CT angiography (CCTA).
METHODS: We randomly assigned 295 patients (168 males, 127 females, median age 71 years) undergoing CCTA to one of two contrast media injection protocols. Group A (n = 150) was injected with a CEO-selected iodine dose based on patient factors. In group B (n = 145), we used our standard protocol (245 mg I/kg). We recorded the CT number in the ascending aorta and determined whether the CT number was equivalent in groups A and B. For the equivalence test, we adopted 75 Hounsfield units (HU) as the equivalence margin. The standard deviation in the CT number and the rate of patients with an acceptable CT number were compared using the F test and the chi-square test, respectively.
RESULTS: The iodine dose in group A was significantly smaller than that in group B (235.7 vs. 253.6 mg I/kg, p < 0.001). The CT number of the ascending aorta was 428.6 ± 55.5 HU in group A and 436.1 ± 68.7 HU in group B; the 95% confidence interval for the difference between the groups was -4.3 HU to 16.9 HU and within the range of the predetermined equivalence margins. In group A, the variance was significantly smaller than that in group B (p = 0.009). The number of patients with an acceptable CT number was significantly higher in group A than in group B (84.7% vs. 71.7%, p = 0.007).
CONCLUSIONS: The use of our CEO for CCTA studies yielded optimal aortic contrast enhancement in significantly more patients than the standard protocol based on the body weight. KEY POINTS: • With our contrast enhancement optimizer (CEO) software, optimal and stable aortic enhancement can be obtained on coronary CT angiography scans irrespective of patient factors. • Management of contrast media becomes more appropriate by the CEO software. • The CEO software can control contrast enhancement at different tube voltage levels.

Entities:  

Keywords:  Cardiac imaging techniques; Cardiac output; Computed tomography angiography; Contrast media; Diagnostic imaging

Mesh:

Substances:

Year:  2018        PMID: 30421021     DOI: 10.1007/s00330-018-5823-2

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


  34 in total

1.  Non-invasive 16-row multislice CT coronary angiography: usefulness of saline chaser.

Authors:  Filippo Cademartiri; Nico Mollet; Aad van der Lugt; Koen Nieman; Peter M T Pattynama; Pim J de Feyter; Gabriel P Krestin
Journal:  Eur Radiol       Date:  2003-12-19       Impact factor: 5.315

2.  Uniform vascular contrast enhancement and reduced contrast medium volume achieved by using exponentially decelerated contrast material injection method.

Authors:  Kyongtae T Bae; Huy Q Tran; Jay P Heiken
Journal:  Radiology       Date:  2004-06       Impact factor: 11.105

Review 3.  Use of high-concentration contrast media in multiple-detector-row CT: principles and rationale.

Authors:  Dominik Fleischmann
Journal:  Eur Radiol       Date:  2003-12       Impact factor: 5.315

4.  Statistical issues in analysis of diagnostic imaging experiments with multiple observations per patient.

Authors:  M Gönen; K S Panageas; S M Larson
Journal:  Radiology       Date:  2001-12       Impact factor: 11.105

5.  Higher intracoronary attenuation improves diagnostic accuracy in MDCT coronary angiography.

Authors:  Filippo Cademartiri; Nico R Mollet; Pedro A Lemos; Fransesco Saia; Massimo Midiri; Pim J de Feyter; Gabriel P Krestin
Journal:  AJR Am J Roentgenol       Date:  2006-10       Impact factor: 3.959

6.  Improved uniformity of aortic enhancement with customized contrast medium injection protocols at CT angiography.

Authors:  D Fleischmann; G D Rubin; A A Bankier; K Hittmair
Journal:  Radiology       Date:  2000-02       Impact factor: 11.105

7.  Intravenous contrast material administration at 16-detector row helical CT coronary angiography: test bolus versus bolus-tracking technique.

Authors:  Filippo Cademartiri; Koen Nieman; Aad van der Lugt; Rolf H Raaijmakers; Nico Mollet; Peter M T Pattynama; Pim J de Feyter; Gabriel P Krestin
Journal:  Radiology       Date:  2004-10-29       Impact factor: 11.105

8.  Effect of contrast material injection duration and rate on aortic peak time and peak enhancement at dynamic CT involving injection protocol with dose tailored to patient weight.

Authors:  Kazuo Awai; Kumiko Hiraishi; Shinichi Hori
Journal:  Radiology       Date:  2004-01       Impact factor: 11.105

9.  Coronary angiography by 64-detector row computed tomography using low dose of contrast material with saline chaser: influence of total injection volume on vessel attenuation.

Authors:  Masaki Yamamuro; Eiji Tadamura; Shotaro Kanao; Yen-Wen Wu; Keiichi Tambara; Masashi Komeda; Masanao Toma; Takeshi Kimura; Toru Kita; Kaori Togashi
Journal:  J Comput Assist Tomogr       Date:  2007 Mar-Apr       Impact factor: 1.826

10.  Optimal dose and injection duration (injection rate) of contrast material for depiction of hypervascular hepatocellular carcinomas by multidetector CT.

Authors:  Yumi Yanaga; Kazuo Awai; Yoshiharu Nakayama; Takeshi Nakaura; Yoshitaka Tamura; Yoshinori Funama; Masahito Aoyama; Naoki Asada; Yasuyuki Yamashita
Journal:  Radiat Med       Date:  2007-07-27
View more
  2 in total

1.  Proposal of a novel protocol using estimated cardiac index fractional dose to improve aortic contrast enhancement for early-phase dynamic CT.

Authors:  Tadashi Kuba; Akihiro Tokushige; Sadayuki Murayama; Shinichiro Ueda
Journal:  Medicine (Baltimore)       Date:  2022-06-24       Impact factor: 1.817

2.  Workload of diagnostic radiologists in the foreseeable future based on recent scientific advances: growth expectations and role of artificial intelligence.

Authors:  Thomas C Kwee; Robert M Kwee
Journal:  Insights Imaging       Date:  2021-06-29
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.