Abstract

Objective: To explore the clinical application value of the spinal robot-assisted surgical system in mild to moderate lumbar spondylolisthesis and evaluate the accuracy of its implantation.
Methods: The clinical data of 56 patients with Meyerding grade Ⅰ or Ⅱ lumbar spondylolisthesis who underwent minimally invasive transforaminal lumbar interbody fusion (MIS-TLIF) between January 2017 and December 2017 were retrospectively analysed. Among them, 28 cases were preoperatively planned with robotic arm and percutaneous pedicle screw placement according to preoperative planning (group A); the other 28 cases underwent fluoroscopy-guided percutaneous pedicle screw placement (group B). There was no significant difference in gender, age, body mass index, slippage type, Meyerding grade, and surgical segmental distribution between the two groups ( P>0.05). The screw insertion angle was measured by CT, the accuracy of screw implantation was evaluated by Neo's criteria, and the invasion of superior articular process was evaluated by Babu's method.
Results: One hundred and twelve screws were implanted in the two groups respectively, 5 screws (4.5%) in group A and 26 screws (23.2%) in group B penetrated the lateral wall of pedicle, and the difference was significant ( χ 2=9.157, P=0.002); the accuracy of nail implantation was assessed according to Neo's criteria, the results were 107 screws of degree 0, 3 of degree 1, 2 of degree 2 in group A, and 86 screws of degree 0, 16 of degree 1, 6 of degree 2, 4 of degree 3 in group B, showing significant difference between the two groups ( Z=4.915, P=0.031). In group B, 20 (17.9%) screws penetrated the superior articular process, while in group A, 80 screws were removed from the decompression side, and only 3 (3.8%) screws penetrated the superior articular process. According to Babu's method, the degree of screw penetration into the facet joint was assessed. The results were 77 screws of grade 0, 2 of grade 1, 1 of grade 2 in group A, and 92 screws of grade 0, 13 of grade 1, 4 of grade 2, 3 of grade 3 in group B, showing significant difference between the two groups ( Z=7.814, P=0.029). The screw insertion angles of groups A and B were (23.5±6.6)° and (18.1±7.5)° respectively, showing significant difference ( t=3.100, P=0.003).
Conclusion: Compared to fluoroscopy-guided percutaneous pedicle screw placement, robot-assisted percutaneous pedicle screw placement has the advantages such as greater accuracy, lower incidence of screw penetration of the pedicle wall and invasion of the facet joints, and has a better screw insertion angle. Combined with MIS-TLIF, robot-assisted percutaneous pedicle screw placement is an effective minimally invasive treatment for lumbar spondylolisthesis.

Entities:  

Keywords:  Lumbar spondylolisthesis; minimally invasive; pedicle screw; spinal robot-assisted surgical system; transforaminal lumbar interbody fusion

Mesh:

Year:  2018        PMID: 30417609     DOI: 10.7507/1002-1892.201804049

Source DB:  PubMed          Journal:  Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi        ISSN: 1002-1892


  6 in total

1.  [Short-term effectiveness comparison between robotic-guided percutaneous minimally invasive pedicle screw internal fixation and traditional open internal fixation in treatment of thoracolumbar fractures].

Authors:  Shu Lin; Jiang Hu; Lun Wan; Liuyi Tang; Yue Wang; Yang Yu; Wei Zhang
Journal:  Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi       Date:  2020-01-15

2.  [A comparative study of spinal robot-assisted and traditional fluoroscopy-assisted percutaneous reduction and internal fixation for single-level thoracolumbar fractures without neurological symptoms].

Authors:  Ye Tian; Jianan Zhang; Hao Chen; Keyuan Ding; Tuanjiang Liu; Dageng Huang; Dingjun Hao
Journal:  Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi       Date:  2020-01-15

3.  [Robot-guided percutaneous kyphoplasty in treatment of multi-segmental osteoporotic vertebral compression fracture].

Authors:  Shu Lin; Jiang Hu; Lun Wan; Liuyi Tang; Yue Wang; Yang Yu; Wei Zhang
Journal:  Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi       Date:  2020-09-15

4.  [Comparative study on effectiveness of modified-transforaminal lumbar interbody fusion and posterior lumbar interbody fusion surgery in treatment of mild to moderate lumbar spondylolisthesis in middle-aged and elderly patients].

Authors:  Zhijie Yang; Bo Liu; Haiyang Lan; He Ye; Jie Chen; Huiqiang Xia; Ye Zhang; Fei Han
Journal:  Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi       Date:  2020-05-15

5.  [A comparative study of robot-assisted and freehand pedicle screw placement in upper thoracic surgery].

Authors:  Fei Wang; Jiang Hu; Liuyi Tang; Lun Wan; Yang Yu; Zhi Tang; Wei Zhang
Journal:  Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi       Date:  2020-12-15

6.  [A comparative study on treatment of lumbar degenerative disease with osteoporosis by manual and robot-assisted cortical bone trajectory screws fixation].

Authors:  Haojie Chen; Shichang Liu; Jianan Zhang; Junsong Yang; Dingjun Hao; Shuai Zhao; Zilong Zhang; Jiarui Yang; Rui Qiao; Xiaoqiang Huang
Journal:  Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi       Date:  2020-09-15
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.