Literature DB >> 30416185

Amperometric Glucose Biosensor Utilizing Zinc Oxide-chitosan-glucose Oxidase Hybrid Composite Films on Electrodeposited Pt-Fe(III).

Eun-Sook Paik1, Yang-Rae Kim2, Hun-Gi Hong1.   

Abstract

We developed an amperometric glucose biosensor based on glucose oxidase (GOx) embedded in zinc oxide (ZnO)-chitosan (CS) hybrid composite films on electrodeposited Pt-Fe(III). This sensor exhibited a fast amperometric response (less than 10 s) to glucose, linearity from 10 μM to 11.0 mM of glucose with a detection limit of 1.0 μM (S/N = 3) and sensitivity of 30.70 μA mM-1 cm-2. An apparent Michaelis-Menten constant of 5.19 mM indicated high affinity between glucose and GOx immobilized in the ZnO-CS films. The effect of interferences such as uric acid, ascorbic acid, and acetaminophen on the performance of this sensor was negligible. In addition, this sensor retained 87% of its initial performance after two weeks of storage at 4°C, indicating that the hybrid composite films allowed successful immobilization of GOx with its high enzymatic activity.

Entities:  

Keywords:  Glucose biosensor; chitosan; composite film; glucose oxidase; zinc oxide

Mesh:

Substances:

Year:  2018        PMID: 30416185     DOI: 10.2116/analsci.18P054

Source DB:  PubMed          Journal:  Anal Sci        ISSN: 0910-6340            Impact factor:   2.081


  1 in total

Review 1.  Exploration of Chitinous Scaffold-Based Interfaces for Glucose Sensing Assemblies.

Authors:  Dipali R Bagal-Kestwal; Been-Huang Chiang
Journal:  Polymers (Basel)       Date:  2019-11-28       Impact factor: 4.329

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.