Is the New Infant Jarvik 2015 Suitable for Patients<8 kg? In Vitro Study Using a Hybrid Simulator.

Arianna Di Molfetta1, Krzysztof Zielinski2, Gianfranco Ferrari2, Macej Kozarski2, Piotr Okrzeja2, Roberta Iacobelli3, Sergio Filippelli3, Gianluigi Perri1, Marek Darowski2, Massimo Massetti1, Robert Jarvik4, Antonio Amodeo3.   

Abstract

Our aim was to study the feasibility of implanting the Infant Jarvik 2015 in patients weighing less than 8 kg. The Infant Jarvik 2015 left ventricular assist device (LVAD) was tested in a hybrid simulator of the cardiovascular system reproducing specific patients' hemodynamics for different patient weights (2-7 kg). For each weight, the sensitivity of the pump to different circulatory parameters (peripheral resistance, left ventricular elastance, right ventricular elastance, heart rate, and heart filling characteristics) has been tested repeating for each experiment a pump ramp (10 000-18 000 rpm). The increase in the pump speed causes a decrease (increase) in the left (right) atrial pressure, an increase (decrease) in the arterial systemic (pulmonary) pressure, an increase in the right ventricular pressure, a decrease (increase) in the left (right) ventricular volume, a decrease in the left ventricular cardiac output, an increase in the LVAD output and an increase in the right ventricular cardiac output (total cardiac output). Suction was observed for lower weight patients and for higher pump speed in the case of vasodilation, left ventricular recovery, bradycardia, right ventricular failure, and left ventricular hypertrophy. Backflow was observed in the case of left ventricular recovery at lower pump speed. In the hybrid simulator, the Infant Jarvik 2015 could be suitable for the implantation in patients lower than 8 kg because of the stability of the device respect to the cardio/circulatory changes (low frequency of suction and backflow) and because of the capability of the device to maintain adequate patient hemodynamics.
© 2018 International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

Entities:  

Keywords:  Hybrid model; Infant Jarvik 2015; Lumped parameter models; Pediatric; Ventricular assist device

Mesh:

Year:  2018        PMID: 30398290     DOI: 10.1111/aor.13302

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  3 in total

1.  Infant miniaturized continuous-flow pumps and permanent support in Pediatrics.

Authors:  Rachele Adorisio; Maria Grandinetti; Antonio Amodeo
Journal:  Ann Cardiothorac Surg       Date:  2021-03

Review 2.  Technology landscape of pediatric mechanical circulatory support devices: A systematic review 2010-2021.

Authors:  Thomas Palazzolo; Matthew Hirschhorn; Ellen Garven; Steven Day; Randy M Stevens; Joseph Rossano; Vakhtang Tchantchaleishvili; Amy L Throckmorton
Journal:  Artif Organs       Date:  2022-04-14       Impact factor: 2.663

Review 3.  Virtual and Artificial Cardiorespiratory Patients in Medicine and Biomedical Engineering.

Authors:  Krzysztof Zieliński; Tomasz Gólczewski; Maciej Kozarski; Marek Darowski
Journal:  Membranes (Basel)       Date:  2022-05-25
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.