Literature DB >> 30396018

Paper-based potentiometric sensing of free bilirubin in blood serum.

Jeffrey G Bell1, Maral P S Mousavi1, Mohamed K Abd El-Rahman2, Edward K W Tan3, Shervanthi Homer-Vanniasinkam4, George M Whitesides5.   

Abstract

Bilirubin is predominantly formed in the liver as a result of breakdown of hemoglobin. Knowing the concentration of bilirubin in serum is important in evaluating the health of the liver, and for the diagnosis of hyperbilirubinemia (a condition that afflicts approximately 60% of full-term and 80% of pre-term newborns). This paper describes the design and fabrication of a potentiometric sensor for the determination of free bilirubin in serum. The sensor has a polymeric ion-selective membrane, and selectively measures free ionic bilirubin ("unbound" bilirubin - i.e., bilirubin not complexed to albumin or other complexing agents), in the presence of other anions - chloride, phosphate, pyruvate, deoxycholate, and lactate - also present in serum. The linear response range of the sensor (1.0 mM to 0.10 μM bilirubin, measured in a sodium phosphate buffer with pH 8.6) covers the clinically-relevant concentration of bilirubin in serum (5-500 μM). Free bilirubin could be detected in human blood serum with this potentiometric sensor. The components of the potentiometric bilirubin sensor were embedded in a paper-based device to provide a sensor that is disposable and easy to use, and thus is suitable for applications at the point-of-care. The paper-based potentiometric bilirubin sensor exhibited a response range of 5.0-0.10 mM (sufficient to cover the clinically-relevant concentration of bilirubin in serum). Only 15 μL of sample is required for measurement of the concentration of free bilirubin, and the analysis can be performed in less than two minutes.
Copyright © 2018 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Free bilirubin; Ion-selective electrode; Jaundice; Liver damage; Paper-based device; Point-of-care diagnostics; Potentiometry

Mesh:

Substances:

Year:  2018        PMID: 30396018     DOI: 10.1016/j.bios.2018.10.055

Source DB:  PubMed          Journal:  Biosens Bioelectron        ISSN: 0956-5663            Impact factor:   10.618


  3 in total

1.  A Novel Method for Measuring Serum Unbound Bilirubin Levels Using Glucose Oxidase-Peroxidase and Bilirubin-Inducible Fluorescent Protein (UnaG): No Influence of Direct Bilirubin.

Authors:  Sota Iwatani; Keiji Yamana; Hajime Nakamura; Kosuke Nishida; Takeshi Morisawa; Masami Mizobuchi; Kayo Osawa; Kazumoto Iijima; Ichiro Morioka
Journal:  Int J Mol Sci       Date:  2020-09-16       Impact factor: 5.923

2.  A stable Ln(iii)-functionalized Cd(ii)-based metal-organic framework: tunable white-light emission and fluorescent probe for monitoring bilirubin.

Authors:  Xiao-Qing Wang; Xuehui Ma; Doudou Feng; Dan Wu; Jie Yang
Journal:  RSC Adv       Date:  2021-11-17       Impact factor: 3.361

3.  Self-Powered Wearable Biosensor in a Baby Diaper for Monitoring Neonatal Jaundice through a Hydrovoltaic-Biosensing Coupling Effect of ZnO Nanoarray.

Authors:  Zirui Ning; Zhihe Long; Guangyou Yang; Lili Xing; Xinyu Xue
Journal:  Biosensors (Basel)       Date:  2022-03-06
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.