Effect of screw access hole design on the fracture resistance of implant-supported zirconia-based restorations.

Abolfazl Saboury1, Minoo Mahshid1, Farhad Tabatabaian1, Leila Moghadam2.   

Abstract

OBJECTIVE: To evaluate the effect of screw access hole design on the fracture resistance of cement-retained implant-supported zirconia-based restorations with screw access holes.
MATERIALS AND METHODS: Thirty cement-retained implant-supported zirconia-based molar crown specimens were fabricated. The specimens were divided into 3 groups of 10: without access holes (C), with screw access holes (S), and with screw access holes accompanied with a surrounding zirconia wall (W). The veneering ceramic thickness was 0.8 mm in all groups. Implants were vertically mounted in a metal block. Abutments were screwed on the implants and the specimens were cemented to the abutments. A universal testing machine applied compressive forces to the specimens until fracture. Fracture resistance values of the specimens were measured. One-way analysis of variance (ANOVA) and Tukey HSD tests were used to analyze data (P < .05).
RESULTS: The mean fracture resistance values were 5794.85 N for C, 2691.48 N for S, and 3878.06 N for W. The screw access hole design significantly affected the fracture resistance (P < .0001). Significant differences were found between C and S (P < .001), C and W (P < .001), and S and W (P = .026).
CONCLUSIONS: The screw access hole decreased the fracture resistance. A surrounding zirconia wall for the screw access hole increased the fracture resistance. CLINICAL SIGNIFICANCE: Screw access hole design may affect the fracture resistance of cement-retained implant-supported zirconia-based restorations with screw access holes. This study introduced a screw access hole design to improve the fracture resistance of these restorations.
© 2018 Wiley Periodicals, Inc.

Entities:  

Keywords:  compressive strength; dental implant; dental prosthesis; fracture resistance; zirconia

Mesh:

Substances:

Year:  2018        PMID: 30394670     DOI: 10.1111/jerd.12422

Source DB:  PubMed          Journal:  J Esthet Restor Dent        ISSN: 1496-4155            Impact factor:   2.843


  3 in total

1.  Esthetic Evaluation and Acceptability of Different Hole Designs on Implant Crowns from the Perspective of Patients and Dentists in China.

Authors:  Cai Wen; Rong Jiang; Zhiqiang Zhang; Bo Lei; Yingquan Zhong; Huangjun Zhou
Journal:  Patient Prefer Adherence       Date:  2022-02-17       Impact factor: 2.711

2.  Selection of 1-mm venting or 2.5-mm screw access holes on implant crowns based on cement extrusion and retention capacity.

Authors:  Huangjun Zhou; Sixian Ye; Min Liu; Hao Feng; Cai Wen
Journal:  BMC Oral Health       Date:  2022-04-02       Impact factor: 2.757

3.  A Special Design to Facilitate Retrieval of Cement-Retained Zirconia-Based Implant-Supported Restorations.

Authors:  Mitra Farzin; Reza Derafshi; Rashin Giti; Masumeh Taghva
Journal:  J Dent (Shiraz)       Date:  2021-06
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.