Removal capacity of different high-flux dialyzers during postdilution online hemodiafiltration.

Alba Santos García1, Nicolás Macías Carmona1, Almudena Vega Martínez1, Soraya Abad Estébanez1, Tania Linares Grávalos1, Inés Aragoncillo Sauco1, Ursula Verdalles Guzmán1, Nayara Panizo González1, Leonidas Cruzado Vega2, Juan Manuel López Gómez1.   

Abstract

INTRODUCTION: The aim of this study is to compare molecule removal and albumin leakage in postdilution online hemodiafiltration with different high-flux dialyzers.
METHODS: We studied seven high-flux dialyzers (Polyflux 210H®, Evodial 2.2®, FxCordiax1000®, Elisio21H®, TS-2.1SL®, XevontaHi20®, VitaPES 210-HF®) in 6 patients. The reduction ratio (RR) of small- and middle-sized molecules was calculated. Dialysate samples were collected to estimate the albumin leakage.
FINDINGS: Global differences between dialyzers were observed in the RR of ß2 microglobulin (P =0.003) and prolactin (P =0.013). The mean loss of albumin in the dialysate per session varied between 114 ± 67 mg (with Evodial 2.2) and 2621 ± 1363 mg per session (with XevontaHi20). We found global differences between dialyzers in total albumin loss (P = 0.05). DISCUSSION: We demonstrated that the performance of high-flux dialyzers was different among the types and that not all high-flux dialyzers should be considered equal.
© 2018 International Society for Hemodialysis.

Entities:  

Keywords:  albumin loss; dialyzer performance; hemodiafiltration; high-flux membrane; middle-sized molecule clearance

Mesh:

Substances:

Year:  2018        PMID: 30367698     DOI: 10.1111/hdi.12685

Source DB:  PubMed          Journal:  Hemodial Int        ISSN: 1492-7535            Impact factor:   1.812


  2 in total

1.  Performance and Hemocompatibility of a Novel Polysulfone Dialyzer: A Randomized Controlled Trial.

Authors:  Götz Ehlerding; Ansgar Erlenkötter; Adelheid Gauly; Bettina Griesshaber; James Kennedy; Lena Rauber; Wolfgang Ries; Hans Schmidt-Gürtler; Manuela Stauss-Grabo; Stephan Wagner; Adam M Zawada; Sebastian Zschätzsch; Manuela Kempkes-Koch
Journal:  Kidney360       Date:  2021-04-07

2.  The removal capacity of asymmetric cellulose triacetate during post-dilution online hemodiafiltration.

Authors:  Alba Santos; Nicolás Macías; Leonidas Cruzado; Almudena Vega
Journal:  Kidney Res Clin Pract       Date:  2021-04-14
  2 in total