Saori Tsukuda1, Nobuhiro Nishii2, Yoshiaki Inoue3, Tatsuya Sasaki4, Hiroshi Morita2, Hiroshi Ito1.
Abstract
Entities:
Keywords: Cardiac resynchronization therapy–defibrillator; Electromagnetic interference; Implantable electronic device; Neurostimulation device; Programming of the neurostimulation devices
Year: 2018 PMID: 30364621 PMCID: PMC6197393 DOI: 10.1016/j.hrcr.2018.06.009
Source DB: PubMed Journal: HeartRhythm Case Rep ISSN: 2214-0271
Figure 1Chest and skull radiographs showing the neurostimulation devices with their electrodes.
Figure 2Postoperative frontal (A) and lateral (B) radiographs showing the leads of the cardiac resynchronization therapy–defibrillator (CRT-D) (black arrows), leads of the neurostimulation devices (NSDs) (black circles), pulse generator of the NSDs (red arrows), and pulse generator of the CRT-D (red circles).
Sensing test and defibrillation test
| Variable | NSDs | CRT-D | ||
|---|---|---|---|---|
| Left device | Right device | RA lead | RV lead | |
| Sensing test of the CRT-D at any output of the NSDs | 60 μs, 130 Hz | 60 μs, 130 Hz | Sensitivity 0.15 ms | Sensitivity 0.15 ms |
| OFF | Bi 2.4 mA | Oversensing (–) | Oversensing (–) | |
| OFF | Bi 3.0 mA | Oversensing (–) | Oversensing (–) | |
| Bi 1.5 mA | OFF | Oversensing (–) | Oversensing (–) | |
| Bi 2.5 mA | OFF | Oversensing (–) | Oversensing (–) | |
| Uni 1.0 mA | Bi 3.0 mA | Oversensing (–) | Oversensing (–) | |
| Uni 2.0 mA | Bi 3.0 mA | Oversensing (–) | Oversensing (–) | |
| Uni 2.5 mA | Bi 3.0 mA | Oversensing (–) | Oversensing (–) | |
| Uni 2.5 mA | Uni 2.0 mA | Oversensing (–) | Oversensing (–) | |
| Uni 2.5 mA | Uni 2.5 mA | Oversensing (–) | Oversensing (–) | |
| Uni 2.5 mA | Uni 3.0 mA | Oversensing (–) | Oversensing (–) | |
| Defibrillation test of the CRT-D | Reset (–) | Reset (–) | ||
When the amplitude of the neurostimulation device (NSD) was set > 3.0 mA on the right stimulator > 2.5 mA on the left stimulator, dyskinesia developed in this patient. Thus, this setting was taken as the maximum tolerable output.
Bi = bipolar setting; CRT-D = cardiac resynchronization therapy–defibrillator; RA = right atrial; RV = right ventricular; Uni = unipolar setting.
No resetting of the NSDs occurred after shock delivery with 20J or 30J.