| Literature DB >> 30338918 |
Richard E Wendt1,2,3, Hua A Ai1,2, Joseph G Meier1,3, Benjamin P Lopez1,3, Samuel J Fahrenholtz1,2, Osama R Mawlawi1,2,3.
Abstract
The attenuation of 511 keV photons by the structure of a PET/MR scanner was measured prior to energizing the magnet. The exposure rate from a source of fluorine-18 was measured in air and, with the source placed at the isocenter of the instrument, at various points outside of the scanner. In an arc from 45 to 135 degrees relative to the long axis of the scanner and at a distance of 1.5 m from the isocenter, the attenuation by the scanner is at least 5.6 half-value layers from the MR component alone and at least 6.6 half-value layers with the PET insert installed. This information could inform better design of the radiation shielding for PET/MR scanners.Entities:
Keywords: PET/MR; self-attenuation
Mesh:
Substances:
Year: 2018 PMID: 30338918 PMCID: PMC6236846 DOI: 10.1002/acm2.12479
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Attenuation of the MR component of the PET/MR scanner. The attenuation values in units of half‐value layer of the MR component of the PET/MR scanner are shown on the polar grid, shown in red, on which the measurements were made. The azimuthal increment of the grid was 22.5 degrees and the radial increment was 0.5 m. The outline of the MR scanner is shown in black.
Half‐value layers of the MR alone
| Radial distance (m) | Azimuthal angle | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 0° | 22.5° | 45° | 67.5° | 90° | 112.5° | 135° | 157.5° | 180° | 225° | 270° | 315° | |
| 1.5 | 0.0 | 0.40 | 5.8 | 7.0 | 6.5 | 7.0 | 5.7 | 0.38 | 0.0 | 5.6 | 6.3 | 6.3 |
| 2 | 0.0 | 0.36 | 5.0 | 6.0 | 6.0 | 6.2 | 4.7 | 0.27 | 0.0 | 4.9 | 6.1 | 5.3 |
| 2.5 | 0.0 | 0.37 | 4.4 | 5.9 | 5.7 | 4.3 | 0.28 | 0.0 | 4.5 | 5.1 | ||
| 3 | 0.0 | 0.39 | 4.4 | 6.4 | 3.8 | 0.29 | 4.1 | 4.2 | ||||
| 3.5 | 0.0 | 0.30 | 4.2 | 0.22 | ||||||||
Figure 2Attenuation of the MR and PET components as well as of the patient handling system of the PET/MR scanner. The attenuation values in units of half‐value layer of the scanner with the PET insert and the patient handling system installed are shown on the polar grid.
Half‐value layers of the MR and PET insert
| Radial distance (m) | Azimuthal angle | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 0° | 22.5° | 45° | 67.5° | 90° | 112.5° | 135° | 157.5° | 180° | 225° | 270° | 315° | |
| 1.5 | 0.0 | 7.2 | 7.8 | 8.0 | 9.1 | 6.6 | 1.9 | 6.9 | 8.1 | 7.5 | ||
| 2 | 0.0 | 0.26 | 5.8 | 7.8 | 7.9 | 8.3 | 5.5 | 1.6 | 5.9 | 7.6 | 6.1 | |
| 2.5 | 0.0 | 0.54 | 5.3 | 6.9 | 7.2 | 5.2 | 1.8 | 1.0 | 5.5 | 5.3 | ||
| 3 | 0.0 | 0.63 | 4.7 | 7.2 | 4.8 | 1.3 | 1.1 | 5.4 | 4.8 | |||
| 3.5 | 0.3 | 0.60 | 4.4 | 0.91 | 0.96 | |||||||
Figure 3Estimated weekly dose map from bare sources simulating a busy work week. The color coding of the dose map is in shades of blue for weekly doses below 20 uSv, which is the limit for the general public in the United States; shades of green for weekly doses between 20 and 100 uSv, which is 10% of the occupational dose limit in the United States and the ALARA 1 limit of the institution in which this scanner was installed; and shades of red for weekly doses of more than 100 uSv, which would exceed the ALARA 1 threshold.
Figure 4Estimated weekly dose map including attenuation by the PET/MR scanner. The red zone is entirely within the scanner room and the green zone does not extend far beyond the walls of the room. The gray areas close to the scanner are very low dose areas (i.e., a “cold” blue). When a fractional occupancy factor for members of the public is applied to the hallway at the top of the figure and to the rooms at the bottom of the figure, there is no need for any structural shielding in this particular example.