Dosimetric effects of brass mesh bolus on skin dose and dose at depth for postmastectomy chest wall irradiation.

Zakiya S Al-Rahbi1, Dean L Cutajar2, Peter Metcalfe3, Anatoly B Rosenfeld4.   

Abstract

PURPOSE: To investigate the feasibility of using the brass mesh bolus as an alternative to tissue- equivalent bolus for post mastectomy chest wall cancer by characterizing the dosimetric effects of the 2-mm fine brass bolus on both the skin dose, the dose at depth and spatial distribution.
MATERIALS AND METHODS: Surface dose and percent depth dose data were acquired for a 6 MV photon beam in a solid water phantom using MOSkin™, Gafchromic EBT3 film and an Advanced Markus ionization chamber. Data were acquired for the case of: no bolus, Face-up bass bolus, Face-down brass bolus, double brass bolus, 0.5 cm and 1.0 cm of Superflab TE bolus. The exit doses were also measured via MOSkin™ dosimeter and Markus ionization chamber. Gafchromic EBT3 film strips were used to plot dose profile at surface and 10 cm depth for Face-up brass, Face-down brass, double brass, 0.5 cm and 1.0 cm of Superflab TE bolus.
RESULTS: The surface dose measured via MOSkin™ dosimeter increased from 19.2 ± 1.0% to 63.1 ± 2.1% under Face-up brass discs, 51.2 ± 1.2% under Face-up brass spaces, 61.5 ± 0.5% under Face-down brass discs, and 41.3 ± 2.1% under Face-down brass spaces. The percentage difference in the dose measured under brass discs between Face-up versus Face-down was less than 2% for entrance dose and 10% for exit dose, whereas the percentage difference under brass spaces was approximately 3% for entrance dose and about 5% for the exit dose. Gafchromic EBT3 film strip measurements show that the mesh bolus produced ripple beam profiles due to the mesh brass construction.
CONCLUSIONS: Brass bolus does not significantly change dose at depth (less than 0.5%), and the surface dose is increased similar to TE bolus. Considering this, brass mesh may be used as a substitute for TE bolus to increase superficial dose for chest wall tangent plans.
Copyright © 2018 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Brass mesh bolus; In vivo skin dosimetry; MOSkin™

Mesh:

Substances:

Year:  2018        PMID: 30337014     DOI: 10.1016/j.ejmp.2018.09.009

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  2 in total

1.  Dosimetric characteristics of a thin bolus made of variable shape tungsten rubber for photon radiotherapy.

Authors:  Katsuya Okuhata; Mikoto Tamura; Hajime Monzen; Yasumasa Nishimura
Journal:  Phys Eng Sci Med       Date:  2021-09-20

2.  Surface Dose Measurements in Chest Wall Postmastectomy Radiotherapy to Achieve Optimal Dose Delivery with 6 MV Photon Beam.

Authors:  Dilson Lobo; Sourjya Banerjee; Challapalli Srinivas; M S Athiyamaan; Shreyas Reddy; Johan Sunny; Ramamoorthy Ravichandran; Himani Kotian; P U Prakash Saxena
Journal:  J Med Phys       Date:  2021-11-20
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.