Natasa Mihailovic1, Cristin Brand2, Larissa Lahme1, Friederike Schubert1, Eike Bormann3, Nicole Eter1, Maged Alnawaiseh1.
Abstract
PURPOSE: To evaluate the repeatability, the reproducibility and the agreement of foveal avascular zone (FAZ) measurements using three different optical coherence tomography angiography (OCT-A) devices. PROCEDURES: This prospective study included 24 eyes of 24 healthy volunteers. OCT-A imaging was performed using RTVue XR Avanti, Canon OCT-HS100 and Spectralis OCT-A. Repeated measurements were performed under the same conditions on two separate days, and the area of the FAZ was determined and analyzed using the above devices.Entities:
Mesh:
Year: 2018 PMID: 30335839 PMCID: PMC6193722 DOI: 10.1371/journal.pone.0206045
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 1A–F. OCT-A images. Representative OCT-A images of the different devices of the SCP (A–C) and DCP (D–F). A+D = Optovue, B+E = Canon, C+F = Heidelberg.
Interobserver (R1 vs. R2) intrasession reproducibility of foveal avascular zone area measurements (m1, m2, m3) of the superficial and deep OCT angiogram for the three different devices.
| SCP | DCP | |
|---|---|---|
| ICC (95% CI) | ICC (95% CI) | |
| m1 | 0.994 (0.987–0.998) | 0.987 (0.969–0.994) |
| m2 | 0.996 (0.990–0.998) | 0.956 (0.898–0.981) |
| m3 | 0.995 (0.988–0.998) | 0.975 (0.943–0.989) |
| m1 | 0.982 (0.959–0.992) | 0.963 (0.917–0.984) |
| m2 | 0.975 (0.942–0.989) | 0.951 (0.891–0.979) |
| m3 | 0.992 (0.982–0.997) | 0.921 (0.826–0.965) |
| m1 | 0.967 (0.926–0.986) | 0.975 (0.942–0.989) |
| m2 | 0.992 (0.982–0.997) | 0.978 (0.951–0.991) |
| m3 | 0.986 (0.968–0.994) | 0.987 (0.969–0.994) |
Pairwise comparison as paired t-test between the first and second (m1 vs. m2) and the first and third measurement (m1 vs. m3) and ICC values for the three different devices (R1, SCP).
| m1: FAZ/mm2 | m2: FAZ/mm2 | m3: FAZ/mm2 | p-value | ICC (95% CI) | |
|---|---|---|---|---|---|
| R1 (m1 vs. m2) | 0.309 ± 0.087 | 0.301 ± 0.090 | 0.308 ± 0.089 | 0.9 | 0.996 (0.991–0.998) |
| R1 (m1 vs. m3) | 0.73 | 0.995 (0.988–0.998) | |||
| R1 (m1 vs. m2) | 0.295 ± 0.093 | 0.295 ± 0.088 | 0.296 ± 0.089 | 0.83 | 0.989 (0.975–0.995) |
| R1 (m1 vs. m3) | 0.70 | 0.987 (0.969–0.994) | |||
| R1 (m1 vs. m2) | 0.326 ± 0.089 | 0.328 ± 0.090 | 0.332 ± 0.097 | 0.67 | 0.994 (0.985–0.997) |
| R1 (m1 vs. m3) | 0.26 | 0.982 (0.960–0.992) |
Pairwise differences and intraclass correlation coefficient (ICC) of FAZ measurements comparing the three different devices (R1, SCP, MD = mean difference, AD = absolute difference).
| Device | FAZ/mm2 | Compared to | MD | AD | p-value | ICC |
|---|---|---|---|---|---|---|
| 0.309 ± 0.087 | 0.01 ± 0.02 | 0.02 ± 0.01 | 0.98 | |||
| -0.02 ± 0.02 | 0.02 ± 0.02 | 0.964 | ||||
| 0.295 ± 0.093 | -0.03 ± 0.03 | 0.04 ± 0.02 | 0.958 | |||
| 0.326 ± 0.089 | - | - | - | - | - |
Fig 2Regression analysis.
Regression line showing the linear relationship between the area of the FAZ measured using the Optovue device and the Canon (blue) / Heidelberg (green) devices.
Fig 3A–C. Bland-Altman-analysis. Bland-Altman plots showing the level of agreement for the area of the FAZ in the superficial retinal OCT-A angiogram. A: Optovue and Canon; B: Optovue and Heidelberg; C: Heidelberg and Canon. Broken line represents the mean difference; dotted lines represent the limits of agreement.