Hydraulic Characterization of Implantable Rotary Blood Pumps.

Stefan Boes, Bente Thamsen, Mattia Haas, Marianne Schmid Daners, Mirko Meboldt, Marcus Granegger.   

Abstract

OBJECTIVE: The hydraulic properties of implantable rotary blood pumps (RBPs) determine their interaction with the cardiovascular system. A systematic comparison in this regard has not yet been performed for different clinically used RBPs. The aim of this study is to describe the hydraulic characteristics of four RBPs with a universal mathematical model and to compare their behavior under clinical operating conditions.
METHODS: First, static and dynamic pump properties of four RBPs (HVAD, Heartmate II, Heartmate 3, and Incor) including their peripheral components were identified in an in vitro setup; results were translated into mathematical models based on principles of turbomachinery including the low and backflow regions. Second, the four hydraulic models were compared in a numerical simulation of the cardiovascular system for full- and partial-support conditions.
RESULTS: A model structure applicable to each of the investigated RBPs was developed. Deviations between simulated and measured signals for static and dynamic properties were small (2.6 ± 0.5 mmHg, 0.38 ± 0.14 L/min, respectively). For a simulated partial support condition, flow pulsatility ranged from 4.1 (Incor) to 9.1 L/min (HVAD). Negative flow rates during diastole were observed in three out of four pumps.
CONCLUSION: Hydraulic properties differ greatly between the investigated RBPs, with flat characteristics for the HVAD and Heartmate II and steeper curves for the Heartmate 3 and especially the Incor. SIGNIFICANCE: Hydraulic characteristics of implantable RBPs are particularly important at lower pump flow rates if backflow is to be avoided. For further research, we provide dynamic hydraulic models of the four RBPs including their periphery.

Entities:  

Mesh:

Year:  2018        PMID: 30334747     DOI: 10.1109/TBME.2018.2876840

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  2 in total

1.  Insights Into the Low Rate of In-Pump Thrombosis With the HeartMate 3: Does the Artificial Pulse Improve Washout?

Authors:  Peng Fang; Jianjun Du; Andrea Boraschi; Silvia Bozzi; Alberto Redaelli; Marianne Schmid Daners; Vartan Kurtcuoglu; Filippo Consolo; Diane de Zélicourt
Journal:  Front Cardiovasc Med       Date:  2022-03-11

2.  Comparison of device-based therapy options for heart failure with preserved ejection fraction: a simulation study.

Authors:  Marcus Granegger; Christoph Gross; David Siemer; Andreas Escher; Sigrid Sandner; Martin Schweiger; Günther Laufer; Daniel Zimpfer
Journal:  Sci Rep       Date:  2022-04-06       Impact factor: 4.379

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.