Precision of Dual-Energy X-Ray Absorptiometry Reflection Scans in Muscular Athletes.

Grant M Tinsley1, Michael Lane Moore2, Austin J Graybeal2.   

Abstract

BACKGROUND: Offset scanning procedures, such as reflection scanning, allow for dual-energy x-ray absorptiometry (DXA) body composition assessment of individuals who are too broad for standard scanning dimensions. However, limited information is available concerning the precision of this procedure, particularly in athletes.
METHODOLOGY: Twenty-seven muscular athletes (n = 17 males, BMI: 28.8 ± 2.0 kg/m2, DXA body fat: 12.5 ± 2.7%; n = 10 females, BMI: 22.8 ± 1.6 kg/m2, DXA body fat: 19.2 ± 3.4%) underwent consecutive DXA scans on a GE Lunar Prodigy scanner using the reflection scanning technique. The fully automated output was obtained for each scan, and an additional version of each scan's output was saved after manual adjustment of regions of interest (ROI). Metrics of reliability and precision were calculated for total and regional body mass (BM), lean mass (LM), fat mass (FM) and bone mineral content (BMC). These metrics included the precision error (PE), least significant change, ΔMean, technical error of measurement, intraclass correlation coefficient, smallest worthwhile effect and minimum difference considered real.
RESULTS: Reflection scanning produced small errors for BM (PE: ∼0.5%), LM (PE < 1%) and BMC (PE: ∼1.2%), with larger errors observed for total FM (PE: ∼3%). Manual ROI adjustment produced lower errors for total BM, LM, FM, and BMC, as well as lower errors for most regional estimates. The utilization of automated ROI revealed concerns unique to reflection scanning, including unnecessary estimation of trunk composition, which appreciably increased error in this region. Regional estimates produced higher errors for all variables as compared to whole-body estimates, although which regions produced the highest errors differed between BM, LM, FM, and BMC.
CONCLUSIONS: Reflection scanning allows DXA body composition assessment in individuals exceeding traditional scanning dimensions, including broad athletes. Although this procedure introduces error, it may be minimized through manual adjustment of ROIs and consistency of analysis methods.
Copyright © 2018 The International Society for Clinical Densitometry. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  DXA; Error; mirrored scanning; offset scanning; reflection scanning

Year:  2018        PMID: 30327241     DOI: 10.1016/j.jocd.2018.09.005

Source DB:  PubMed          Journal:  J Clin Densitom        ISSN: 1094-6950            Impact factor:   2.617


  3 in total

1.  Whole-body dual-energy X-ray absorptiometry demonstrates better reliability than segmental body composition analysis in college-aged students.

Authors:  Petr Kutáč; Václav Bunc; Martin Sigmund
Journal:  PLoS One       Date:  2019-04-22       Impact factor: 3.240

2.  Regional Lean Soft Tissue and Intracellular Water Are Associated with Changes in Lower-Body Neuromuscular Performance: A Pilot Study in Elite Soccer Players.

Authors:  Tindaro Bongiovanni; Grant Tinsley; Giulia Martera; Carmine Orlandi; Federico Genovesi; Giuseppe Puleo; Alessio Rossi; Athos Trecroci
Journal:  Eur J Investig Health Psychol Educ       Date:  2022-07-22

3.  Changes in total and segmental bioelectrical resistance are correlated with whole-body and segmental changes in lean soft tissue following a resistance training intervention.

Authors:  Grant M Tinsley; Patrick S Harty; M Lane Moore; Jozo Grgic; Analiza M Silva; Luis B Sardinha
Journal:  J Int Soc Sports Nutr       Date:  2019-11-29       Impact factor: 5.150

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.