Literature DB >> 30318499

Safety and Efficacy of Transvenous Lead Extraction With a High-Frequency Excimer Laser - A Single Center Experience.

Takakatsu Yoshitake1, Masahiko Goya1, Takeshi Sasaki1, Shinya Shiohira1, Masahiro Sekigawa1, Yasuhiro Shirai1, Kiko Lee1, Atsuhiko Yagishita1, Shingo Maeda2, Yoshihide Takahashi2, Mihoko Kawabata1, Kenzo Hirao1.   

Abstract

BACKGROUND: Transvenous lead extractions have been performed using 40-Hz laser sheaths. Recently, a new 80-Hz laser sheath became available, but only a few reports have compared the effectiveness of the 40- and 80-Hz laser sheaths. Methods and 
Results: This study included 215 patients. Lead extractions using only laser sheaths were analyzed. The clinical characteristics, extraction parameters, and extraction tools were evaluated. The procedures were performed with 40-Hz sheaths in 150 patients (group 1: 270 leads) and 80-Hz sheaths in 65 (group 2: 99 leads). No statistically significant differences were observed in the clinical parameters except for sex. The mean implant duration was 95.3±86.0 and 78.2±56.8 months in groups 1 and 2, respectively (P=0.07). The respective mean laser time and number of laser pulses were 48.5±52.1 and 48.1±56.1 s (P=0.96) and 2,035.0±2,384.0 and 3,955.1±2,339.3 pulses (P<0.0001). Complete removal was achieved for 97.4% of the leads and in 98.0% in both groups (P=0.38). Major complications occurred in 2.0% and 3.1% of the patients in groups 1 and 2, respectively (P=0.94).
CONCLUSIONS: Transvenous lead extraction using high-frequency laser sheaths was as highly successful as with low-frequency laser sheaths in Japanese patients.

Entities:  

Keywords:  High-frequency laser sheath; Implantable cardioverter-defibrillator (ICD) lead; Laser sheath; Lead extraction; Pacemaker lead

Mesh:

Year:  2018        PMID: 30318499     DOI: 10.1253/circj.CJ-18-0869

Source DB:  PubMed          Journal:  Circ J        ISSN: 1346-9843            Impact factor:   2.993


  1 in total

1.  Results of the Patient-Related Outcomes of Mechanical lead Extraction Techniques (PROMET) study: a multicentre retrospective study on advanced mechanical lead extraction techniques.

Authors:  Christoph T Starck; Elkin Gonzalez; Omar Al-Razzo; Patrizio Mazzone; Peter-Paul Delnoy; Alexander Breitenstein; Jan Steffel; Jürgen Eulert-Grehn; Pia Lanmüller; Francesco Melillo; Alessandra Marzi; Manav Sohal; Giulia Domenichini; Mark M Gallagher
Journal:  Europace       Date:  2020-07-01       Impact factor: 5.214