Cryobiopsy in the Upper Urinary Tract: Preclinical Evaluation of a Novel Device.

Jan-Thorsten Klein1, Franziska Berger1, Walter Linzenbold2, Luise Jäger2, Markus D Enderle2, Hans Bösmüller3, Jens Mundhenk4, Christian Schwentner4, Christian Bolenz5.   

Abstract

OBJECTIVE: To develop a novel device for cryobiopsy of the upper urinary tract (UUT) and to evaluate its feasibility in a standardized preclinical setting.
MATERIALS AND METHODS: Flexible cryoprobes (diameter 0.9 mm; cooling agent CO2) were developed and used to extract biopsies in porcine UUTs. Cryosamples obtained by ureterorenoscopy were systematically compared with biopsy specimens obtained with standard of care devices in terms of physical characteristics (deflection angle and irrigation flow rates) and histologic criteria (assessability).
RESULTS: Irrigation flow rates were significantly higher with introduced BIGopsy (2.8 ± 0.1) compared with standard forceps (0.94 ± 0.06; P < .001) and cryoprobe (1.1 ± 0.1; P < .001). Angular deflection was significantly reduced by the inserted cryoprobe (130.7° ± 1.2° vs 166.9° ± 1.1° [BIGopsy] or 161.4° ± 1.9° [standard forceps]; both P < .001). Significantly larger UUT tissue samples were obtained by the cryoprobe (mean specimen area 7.5 ± 2.5 vs 4.6 ± 2.5 mm² [BIGopsy] or 1.4 ± 1.4 mm² [standard forceps]; both P < .001). No crush artifacts were observed in cryosamples. Superior histologic assessability scores were achieved in samples obtained by the cryoprobe (mean 2.8 ± 0.8) and BIGopsy (2.3 ± 1.9) when compared with standard forceps (0.4 ± 0.9; P < .001).
CONCLUSION: Cryobiopsy in the UUT is feasible and represents a viable new option to improve the diagnostic accuracy of histopathologic evaluation. Larger and more representative tissue samples can be obtained using a cryoprobe and artifacts may be avoided. Further optimization of the probe will reduce possible restrictions of ureterorenoscopy handling when the device is inserted.
Copyright © 2018. Published by Elsevier Inc.

Entities:  

Mesh:

Year:  2018        PMID: 30312669     DOI: 10.1016/j.urology.2018.10.003

Source DB:  PubMed          Journal:  Urology        ISSN: 0090-4295            Impact factor:   2.649


  2 in total

1.  A new tool for bile duct tissue sampling: ex vivo clinical evaluation of intraductal cryobiopsy for cholangioscopy.

Authors:  Lukas Wirsing; Walter Linzenbold; Simon U Jaeger; Phillip Stahl; German Ott; Tobias Leibold; Markus Enderle; Jörg Albert; Jan Peveling-Oberhag
Journal:  Endosc Int Open       Date:  2022-06-10

2.  Improving the Quality of Human Upper Urinary Tract Specimens by Cryobiopsy.

Authors:  Jan T Klein; Axel John; Lars Bohnert; Markus D Enderle; Walter Linzenbold; Christian Bolenz
Journal:  Front Oncol       Date:  2022-02-11       Impact factor: 6.244

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.