Usefulness of slice encoding for metal artifact correction (SEMAC) technique for reducing metal artifacts after total knee arthroplasty.

Ahmed Jawhar1, Miriam Reichert2, Michael Kostrzewa3, Mathias Nittka4, Ulrike Attenberger2, Henning Roehl5, Frederic Bludau5.   

Abstract

PURPOSE: To evaluate the usefulness of a novel MRI sequence strategy in the assessment of the periprosthetic anatomical structures after primary total knee arthroplasty.
METHODS: Two MR sequences were retrospectively compared for the imaging of 15 patients with implanted cruciate-retaining/fixed-bearing TKAs (DePuy, PFC Sigma): a slice encoding sequence for metal artifact correction (SEMAC) and a standard sequence. Images were acquired on a 1.5-T system. The degree of artifact reduction was assessed using several qualitative (Likert-type scale) (artifact size, distorsion, blur, image quality, periprosthetic bone, posterior cruciate ligament, lateral collateral ligament, medial collateral ligament, patella tendon, popliteal vessels) and quantitative (artifact volume, Insall-Salvati index, length of patella/tendon, prosthesis dimensions) parameters by blinded reads performed by four investigators. The SEMAC sequences were statistically compared with the standard sequence using Wilcoxon test. Additionally, the intraclass correlation coefficient (ICC) for interobserver agreement was calculated.
RESULTS: Higher levels of blurring were found with SEMAC compared to standard sequences (p < 0.001). All other qualitative parameters improved significantly with the application of SEMAC. In comparison with conventional sequences, the artifact volume was reduced by 59% utilizing SEMAC. Thus, the artifact reduction improved the precision of measurements such as Insall-Salvati index and length of patella/tendon (p < 0.001). The dimension of the tibial component (Ti alloy/polyethylene) revealed accurate values with both MRI sequences. A sufficient interobserver agreement among all readers was found with SEMAC, qualitatively ICC 0.9 (range 0.8-1) as well as quantitatively ICC 0.95 (range 0.92-0.98).
CONCLUSIONS: SEMAC effectively reduces artifacts caused by metallic implants after total knee arthroplasty relative to standard imaging. This allows for an improved assessment of periprosthetic anatomical structures. This might enable an improved detectability of postoperative complications in the future. LEVEL OF EVIDENCE: Diagnostic Study Level III.

Entities:  

Keywords:  Artifact reduction; Ligaments; Periprosthetic bone; SEMAC; Total knee arthroplasty

Mesh:

Substances:

Year:  2018        PMID: 30310989     DOI: 10.1007/s00590-018-2322-8

Source DB:  PubMed          Journal:  Eur J Orthop Surg Traumatol        ISSN: 1633-8065


  22 in total

1.  Quantitative assessment of an MR technique for reducing metal artifact: application to spin-echo imaging in a phantom.

Authors:  M J Lee; D L Janzen; P L Munk; A MacKay; Q S Xiang; A McGowen
Journal:  Skeletal Radiol       Date:  2001-07       Impact factor: 2.199

2.  Tibial component failure mechanisms in total knee arthroplasty.

Authors:  Michael E Berend; Merrill A Ritter; John B Meding; Philip M Faris; E Michael Keating; Ryan Redelman; Gregory W Faris; Kenneth E Davis
Journal:  Clin Orthop Relat Res       Date:  2004-11       Impact factor: 4.176

3.  Alteration of the patellar height following total knee arthroplasty.

Authors:  A Jawhar; S Sohoni; V Shah; H P Scharf
Journal:  Arch Orthop Trauma Surg       Date:  2013-11-28       Impact factor: 3.067

4.  Is the effect of a posterior cruciate ligament resection in total knee arthroplasty predictable?

Authors:  Christoph Schnurr; Peer Eysel; Dietmar P König
Journal:  Int Orthop       Date:  2011-06-25       Impact factor: 3.075

5.  Detection of periprosthetic osteolysis around total knee arthroplasties an in vitro study.

Authors:  Mario G T Zotti; David G Campbell; Richard Woodman
Journal:  J Arthroplasty       Date:  2012-02       Impact factor: 4.757

6.  Exploration of radiographically normal total knee replacements for unexplained pain.

Authors:  M A Mont; F K Serna; K A Krackow; D S Hungerford
Journal:  Clin Orthop Relat Res       Date:  1996-10       Impact factor: 4.176

7.  Late rupture of the posterior cruciate ligament after total knee replacement.

Authors:  R L Montgomery; S B Goodman; J Csongradi
Journal:  Iowa Orthop J       Date:  1993

8.  The effect of the Insall-Salvati ratio on outcome after total knee arthroplasty.

Authors:  R Michael Meneghini; Merrill A Ritter; Jeffery L Pierson; John B Meding; Michael E Berend; Philip M Faris
Journal:  J Arthroplasty       Date:  2006-09       Impact factor: 4.757

9.  Metal artefact reduction in MRI at both 1.5 and 3.0 T using slice encoding for metal artefact correction and view angle tilting.

Authors:  M Reichert; T Ai; J N Morelli; M Nittka; U Attenberger; V M Runge
Journal:  Br J Radiol       Date:  2015-01-23       Impact factor: 3.039

10.  SEMAC: Slice Encoding for Metal Artifact Correction in MRI.

Authors:  Wenmiao Lu; Kim Butts Pauly; Garry E Gold; John M Pauly; Brian A Hargreaves
Journal:  Magn Reson Med       Date:  2009-07       Impact factor: 4.668

View more
  2 in total

1.  MRI as Diagnostic Modality for Analyzing the Problematic Knee Arthroplasty: A Systematic Review.

Authors:  Femke F Schröder; Corine E Post; Frank-Christiaan B M Wagenaar; Nico Verdonschot; Rianne M H A Huis In't Veld
Journal:  J Magn Reson Imaging       Date:  2019-07-22       Impact factor: 4.813

2.  Quantification of intra-articular fibrosis in patients with stiff knee arthroplasties using metal-reduction MRI.

Authors:  Veronica Attard; Cheuk Yin Li; Alexander Self; Derek A Mann; Lee A Borthwick; Phil O'Connor; David J Deehan; Nicholas S Kalson
Journal:  Bone Joint J       Date:  2020-10       Impact factor: 5.082

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.