PURPOSE: Achieving an effective seal with the Nellix endovascular aneurysm system is dependent on filling the stent bags with an appropriate volume of polymer. Calculating this volume preoperatively is essential and can be performed manually or using three-dimensional (3D) software. The aim of this project was to compare the accuracy of these two methods relative to the actual polymer volume used.Entities:
Keywords: Computed tomography; Endovascular sealing; Nellix graft; Polymers; Three-dimensional
Year: 2018 PMID: 30310808 PMCID: PMC6175572 DOI: 10.5758/vsi.2018.34.3.61
Source DB: PubMed Journal: Vasc Specialist Int ISSN: 2288-7970
Fig. 1Manual method for calculating Nellix polymer volume.
Fig. 2Three-dimensional-derived method for calculating Nellix polymer volume.
Demographics, AAA anatomy and device type
| Value | Patient (n=28) |
|---|---|
| Age (y) | 80.9±5.596 |
| Male | 26 (92.9) |
| Aortic aneurysm neck length | |
| <15 mm | 2 (7.1) |
| >15 mm | 26 (92.9) |
| Asymptomatic AAA | 26 (92.9) |
| Symptomatic AAA | 1 (3.6) |
| Rupture AAA | 1 (3.6) |
| Type of repair | |
| EVAS | 26 (92.9) |
| Ch-EVAS | 2 (7.1) |
Values are presented as mean±standard deviation or number (%). AAA, abdominal aortic aneurysm; EVAS, endovascular aneurysm sealing; Ch-EVAS, chimney endovascular aneurysm sealing.
Difference between manual and 3D volume in comparison to real volume
| Minimum | Maximum | Mean | SD | |
|---|---|---|---|---|
| Age (y) | 70 | 92 | 80.86 | 5.596 |
| Manual volume (mL) | 47 | 186 | 100.14 | 37.115 |
| 3D volume (mL) | 52 | 225 | 110.18 | 45.560 |
| Real volume (mL) | 34 | 214 | 103.00 | 46.600 |
| Manual volume % difference (mL) | −29 | 41 | 1.11 | 15.643 |
| 3D volume % difference (mL) | −8 | 53 | 9.48 | 11.501 |
3D, three-dimensional; SD, standard deviation.
Paired samples test
| Paired differences | t | df | P-value | ||||||
|---|---|---|---|---|---|---|---|---|---|
|
| |||||||||
| Mean | Standard deviation | Standard error mean | 95% confidence interval-difference | ||||||
|
| |||||||||
| Upper | Lower | ||||||||
| Pair 1 | Manual volume-Real volume | −2.857 | 16.436 | 3.106 | −9.230 | 3.516 | −0.920 | 27 | 0.366 |
| Pair 2 | Three-dimensional volume-Real volume | 7.179 | 7.399 | 1.398 | 4.310 | 10.048 | 5.134 | 27 | 0.000 |