In Vitro Flow Rate Dependency of Delivered Dose and Fine Particle Dose of Salmeterol/Fluticasone Propionate Easyhaler and Seretide Diskus with Patient Flow Rates Collected in a Randomized Controlled Trial.

Rain Jõgi1, Satu Lähelmä2, Mikko Vahteristo2, Anita Happonen2, Jussi Haikarainen3.   

Abstract

BACKGROUND: The Easyhaler® device-metered dry powder inhaler containing Salmeterol and Fluticasone propionate (S/F) has been developed for the treatment of patients with asthma and chronic obstructive pulmonary disease (COPD). We report two studies which evaluated the in vitro flow rate dependence of delivered dose (DD) and fine particle dose (FPD) of S/F Easyhaler versus Seretide Diskus®.
METHODS: A randomized controlled trial (RCT) assessed inspiratory flow parameters of S/F Easyhaler and Seretide Diskus in subgroups of patients with asthma (children, adolescents and adults, and elderly) and in COPD patients. The 10th, 50th, and 90th percentile airflow rates were determined and utilized in vitro, to evaluate flow rate dependence of DD and FPD. Flow rate dependence was evaluated relative to the result obtained at the 50th percentile and any values deviating from 100% indicated flow rate dependence. The volumetric flow rate dependence (Q) index derived from FPD at 10th and 90th percentile airflows was also evaluated.
RESULTS: Overall, 227 patients were enrolled and randomized; 216 completed the RCT. In total, 55.5% of patients were female, and the mean age was 46.3 years. Clinically relevant airflow rates (46, 68, and 85 L/min for S/F Easyhaler and 44, 71, and 96 L/min for Seretide Diskus) were carried forward into the in vitro study, which demonstrated similar flow rate dependence of DD and FPD for S/F Easyhaler compared with Seretide Diskus; all values were within ±15% limits across the 10th, 50th, and 90th percentile airflow rates. Q index results suggested that both S/F Easyhaler and Seretide Diskus are medium airflow-dependent products.
CONCLUSIONS: Similar in vitro flow rate dependence of DD and FPD was demonstrated for S/F Easyhaler compared with Seretide Diskus, across a range of clinically relevant airflow rates, collected from patients with asthma and COPD.

Entities:  

Keywords:  Salmeterol/Fluticasone Easyhaler; delivered dose; fine particle dose; flow rate dependency

Year:  2018        PMID: 30300557      PMCID: PMC6477585          DOI: 10.1089/jamp.2018.1463

Source DB:  PubMed          Journal:  J Aerosol Med Pulm Drug Deliv        ISSN: 1941-2711            Impact factor:   2.849


  22 in total

Review 1.  Idealhalers or realhalers? A comparison of Diskus and Turbuhaler.

Authors:  L Borgström; L Asking; L Thorsson
Journal:  Int J Clin Pract       Date:  2005-12       Impact factor: 2.503

2.  Peak inspiratory flow through Turbuhaler in acute asthma.

Authors:  P H Brown; A C Ning; A P Greening; A McLean; G K Crompton
Journal:  Eur Respir J       Date:  1995-11       Impact factor: 16.671

3.  The relationship between powder inhaler resistance and peak inspiratory conditions in healthy volunteers--implications for in vitro testing.

Authors:  A R Clark; A M Hollingworth
Journal:  J Aerosol Med       Date:  1993

4.  Novolizer: how does it fit into inhalation therapy?

Authors:  Helgo Magnussen
Journal:  Curr Med Res Opin       Date:  2005       Impact factor: 2.580

5.  Inspiratory flows through dry powder inhaler in chronic obstructive pulmonary disease: age and gender rather than severity matters.

Authors:  L Pekka Malmberg; Paula Rytilä; Pertti Happonen; Tari Haahtela
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2010-08-09

6.  [Patient preference in the choice of dry powder inhalers].

Authors:  J Giner; M Torrejón; A Ramos; P Casan; C Granel; V Plaza; J Belda; J Sanchis
Journal:  Arch Bronconeumol       Date:  2004-03       Impact factor: 4.872

7.  Evaluation of in vitro and in vivo flow rate dependency of budesonide/formoterol Easyhaler(®).

Authors:  L Pekka Malmberg; Mark L Everard; Jussi Haikarainen; Satu Lähelmä
Journal:  J Aerosol Med Pulm Drug Deliv       Date:  2014-06-30       Impact factor: 2.849

Review 8.  Critical inhaler errors in asthma and COPD: a systematic review of impact on health outcomes.

Authors:  Omar Sharif Usmani; Federico Lavorini; Jonathan Marshall; William Christopher Nigel Dunlop; Louise Heron; Emily Farrington; Richard Dekhuijzen
Journal:  Respir Res       Date:  2018-01-16

Review 9.  Device errors in asthma and COPD: systematic literature review and meta-analysis.

Authors:  Henry Chrystyn; Job van der Palen; Raj Sharma; Neil Barnes; Bruno Delafont; Anadi Mahajan; Mike Thomas
Journal:  NPJ Prim Care Respir Med       Date:  2017-04-03       Impact factor: 2.871

10.  Pharmacokinetics of Salmeterol and Fluticasone Propionate Delivered in Combination via Easyhaler and Diskus Dry Powder Inhalers in Healthy Subjects.

Authors:  Merja Kirjavainen; Leena Mattila; Mikko Vahteristo; Jani Korhonen; Satu Lähelmä
Journal:  J Aerosol Med Pulm Drug Deliv       Date:  2018-03-01       Impact factor: 2.849

View more
  7 in total

Review 1.  Understanding Dry Powder Inhalers: Key Technical and Patient Preference Attributes.

Authors:  Mark L Levy; Will Carroll; José L Izquierdo Alonso; Claus Keller; Federico Lavorini; Lauri Lehtimäki
Journal:  Adv Ther       Date:  2019-09-02       Impact factor: 3.845

2.  Patients with asthma or chronic obstructive pulmonary disease (COPD) can generate sufficient inspiratory flows via Easyhaler® dry powder inhaler: a pooled analysis of two randomized controlled trials.

Authors:  L Pekka Malmberg; Anna S Pelkonen; Ville Vartiainen; Mikko Vahteristo; Satu Lähelmä; Rain Jõgi
Journal:  J Thorac Dis       Date:  2021-02       Impact factor: 3.005

Review 3.  Lung Deposition and Inspiratory Flow Rate in Patients with Chronic Obstructive Pulmonary Disease Using Different Inhalation Devices: A Systematic Literature Review and Expert Opinion.

Authors:  Adolfo Baloira; Araceli Abad; Antonia Fuster; Juan Luis García Rivero; Patricia García-Sidro; Eduardo Márquez-Martín; Marta Palop; Néstor Soler; J L Velasco; Fernando González-Torralba
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2021-04-19

4.  Inspiratory Flow Parameters Through Dry Powder Inhalers in Healthy Volunteers and Patients with Chronic Obstructive Pulmonary Disease (COPD): Device Resistance Does Not Limit Use in COPD.

Authors:  Rain Jõgi; Leena Mattila; Mikko Vahteristo; Aino Takala; Satu Lähelmä; Ville A Vartiainen; Ari Lindqvist
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2021-04-30

Review 5.  Relationship between Peak Inspiratory Flow and Patient and Disease Characteristics in Individuals with COPD-A Systematic Scoping Review.

Authors:  Marika T Leving; Janwillem Kocks; Sinthia Bosnic-Anticevich; Richard Dekhuijzen; Omar S Usmani
Journal:  Biomedicines       Date:  2022-02-16

6.  Characterization and comparison of Re-Du-Ning aerosol particles generated by different jet nebulizers.

Authors:  Yi Zhang; Brian Chi-Yan Cheng; Cui Li; Yonghua Tao; Chanjuan Yu; Xinyue Liu; Xiaoyan Gao; Gan Luo
Journal:  RSC Adv       Date:  2019-09-25       Impact factor: 4.036

7.  A non-interventional switch study in adult patients with asthma or COPD on clinical effectiveness of salmeterol/fluticasone Easyhaler® in routine clinical practice.

Authors:  Ines Vinge; Jörgen Syk; Athanasios Xanthopoulos; Hendrik Laßmann; Mikko Vahteristo; Ulla Sairanen; Satu Lähelmä; Rudolf Hennig; Matthias Müller
Journal:  Ther Adv Respir Dis       Date:  2021 Jan-Dec       Impact factor: 4.031

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.