PURPOSE: To report the outcomes of LASIK for mixed cylinder using the Triple-A ablation profile with the MEL 90 excimer laser (Carl Zeiss Meditec, Jena, Germany). METHODS: This was a retrospective analysis of all eyes treated by LASIK for mixed cylinder using the Triple-A ablation profile with the MEL 90 laser between July 2013 and October 2016. Patients were observed for 1 year after surgery. Standard outcomes analysis and vector analysis by the Alpins method were performed. RESULTS: The database review identified 105 eyes (82 patients) treated by LASIK for mixed cylinder using the MEL 90 laser. Mean age was 40 ± 11 years (range: 18 to 65 years). Mean attempted spherical equivalent was +0.30 ± 0.90 diopters (D) (range: -2.30 to +1.75 D). Mean cylinder was -2.93 ± 1.47 D (range: -0.75 to -7.00 D). Preoperative corrected distance visual acuity (CDVA) was 20/20 or better in 81% of eyes. Postoperative uncorrected distance visual acuity was 20/20 or better in 69% and 20/25 or better in 86% of eyes. Mean postoperative spherical equivalent relative to the intended target was -0.21 ± 0.38 D (range: -1.25 to +1.13 D), with 83% within ±0.50 D. Mean postoperative cylinder was -0.57 ± 0.41 D (range: 0.00 to -1.75 D). Geometric mean was 1.12 for the correction index and 0.25 for the index of success. For angle of error, the arithmetic mean was -0.6° ± 4.2° (range: -13.5° to 9.5°) and the absolute mean was 2.9° ± 3.0° (range: 0.0° to 13.5°). There was loss of one line of CDVA in 3% of eyes and no eyes lost two or more lines CDVA. There was a small increase in contrast sensitivity after surgery at 3, 6, 12, and 18 cycles per degree (P < .05). CONCLUSIONS: One-year outcomes of LASIK using the MEL 90 laser for mixed cylinder up to -7.00 D demonstrated excellent efficacy, safety, and predictability. Vector analysis found a 12% overcorrection in magnitude of refractive cylinder. Given the high accuracy for angle of error, a nomogram could be applied to reduce the over-correction and further improve the uncorrected distance visual acuity outcomes. [J Refract Surg. 2018;34(10):672-680.]. Copyright 2018, SLACK Incorporated.
PURPOSE: To report the outcomes of LASIK for mixed cylinder using the Triple-A ablation profile with the MEL 90 excimer laser (Carl Zeiss Meditec, Jena, Germany). METHODS: This was a retrospective analysis of all eyes treated by LASIK for mixed cylinder using the Triple-A ablation profile with the MEL 90 laser between July 2013 and October 2016. Patients were observed for 1 year after surgery. Standard outcomes analysis and vector analysis by the Alpins method were performed. RESULTS: The database review identified 105 eyes (82 patients) treated by LASIK for mixed cylinder using the MEL 90 laser. Mean age was 40 ± 11 years (range: 18 to 65 years). Mean attempted spherical equivalent was +0.30 ± 0.90 diopters (D) (range: -2.30 to +1.75 D). Mean cylinder was -2.93 ± 1.47 D (range: -0.75 to -7.00 D). Preoperative corrected distance visual acuity (CDVA) was 20/20 or better in 81% of eyes. Postoperative uncorrected distance visual acuity was 20/20 or better in 69% and 20/25 or better in 86% of eyes. Mean postoperative spherical equivalent relative to the intended target was -0.21 ± 0.38 D (range: -1.25 to +1.13 D), with 83% within ±0.50 D. Mean postoperative cylinder was -0.57 ± 0.41 D (range: 0.00 to -1.75 D). Geometric mean was 1.12 for the correction index and 0.25 for the index of success. For angle of error, the arithmetic mean was -0.6° ± 4.2° (range: -13.5° to 9.5°) and the absolute mean was 2.9° ± 3.0° (range: 0.0° to 13.5°). There was loss of one line of CDVA in 3% of eyes and no eyes lost two or more lines CDVA. There was a small increase in contrast sensitivity after surgery at 3, 6, 12, and 18 cycles per degree (P < .05). CONCLUSIONS: One-year outcomes of LASIK using the MEL 90 laser for mixed cylinder up to -7.00 D demonstrated excellent efficacy, safety, and predictability. Vector analysis found a 12% overcorrection in magnitude of refractive cylinder. Given the high accuracy for angle of error, a nomogram could be applied to reduce the over-correction and further improve the uncorrected distance visual acuity outcomes. [J Refract Surg. 2018;34(10):672-680.]. Copyright 2018, SLACK Incorporated.