Amy K Saenger1, Allan S Jaffe2, Richard Body3,4,5, Paul O Collinson6, Peter A Kavsak7, Carolyn S P Lam8, Guillaume Lefèvre9, Tobjørn Omland10,11, Jordi Ordóñez-Llanos12,13, Kari Pulkki14, Fred S Apple15,16.
Abstract
Two interferences recently brought to the forefront as patient safety issues include hemolysis (hemoglobin) and biotin (vitamin B7). The International Federation for Clinical Chemistry Committee on Cardiac Biomarkers (IFCC-CB) obtained input from a majority of cTn and NP assay manufacturers to collate information related to high-sensitivity (hs)-cTnI, hs-cTnT, contemporary, and POC cTn assays, and NP assays interferences due to hemolysis and biotin. The information contained in these tables was designed as educational tools to aid laboratory professionals and clinicians in troubleshooting cardiac biomarker analytical results that are discordant with the clinical situation.Entities:
Keywords: biotin; hemolysis; interferences; natriuretic peptide; troponin