Sepiso K Masenga1,2, Mercy Mweemba3, Annie Kachele4, Yahns Chalubemba4, Paul Toloka4, Benson M Hamooya4,5.
Abstract
OBJECTIVE: To synthesize lessons from comparison of results obtained from the immuno-chromatographic SD Bioline testing method and the chemiluminescent micro-particle immunoassay Architect in the diagnosis of syphilis at Livingstone Central hospital laboratory.Entities:
Keywords: Architect syphilis; Immunoassay; SD Bioline; Treponema pallidum
Mesh:
Substances:
Year: 2018 PMID: 30286790 PMCID: PMC6172779 DOI: 10.1186/s13104-018-3808-5
Source DB: PubMed Journal: BMC Res Notes ISSN: 1756-0500
Positivity rate of suspected syphilis samples using SD Bioline test
| SD Bioline test | March | April | May | Total |
|---|---|---|---|---|
| Positive | 44 (19.8%) | 46 (21.5%) | 26 (11.4%) | 116 (14.9%) |
| Negative | 222 (80.2%) | 214 (78.5%) | 229 (88.6%) | 665 (85.1%) |
| Total | 266 (100%) | 260 (100%) | 255 (100%) | 781 (100%) |
False positivity and false negative rates of SD Bioline compared with architect syphilis test
| Architect TP | ||||
|---|---|---|---|---|
| Positive | Negative | |||
| SD Bioline | Positive | 42 (91.3%) | 8 (14.8) | < 0.0005 |
| Negative | 4 (8.7%) | 46 (85.2%) | ||
| Total | 46 (100%) | 54 (100%) | ||
p-value, Fishers’ exact test used
Cohen’s kappa (ĸ) results showing a good degree of concordance between SD Bioline and Architect tests
| Agreement | ĸ | Standard error | p value |
|---|---|---|---|
| 88% | 0.760 | 0.0997 | p < 0.0005 |