| Literature DB >> 30284642 |
Marly van Assen1, Gert Jan Pelgrim1, Emmy Slager1, Sjoerd van Tuijl2, U Joseph Schoepf3, Rozemarijn Vliegenthart4, Matthijs Oudkerk5.
Abstract
The purpose of this study was to evaluate the effect of temporal sampling rate in dynamic CT myocardial perfusion imaging (CTMPI) on myocardial blood flow (MBF). Dynamic perfusion CT underestimates myocardial blood flow compared to PET and SPECT values. For accurate quantitative analysis of myocardial perfusion with dynamic perfusion CT a stable calibrated HU measurement of MBF is essential. Three porcine hearts were perfused using an ex-vivo Langendorff model. Hemodynamic parameters were monitored. Dynamic CTMPI was performed using third generation dual source CT at 70 kVp and 230-350 mAs/rot in electrocardiography(ECG)-triggered shuttle-mode (sampling rate, 1 acquisition every 2-3 s; z-range, 10.2 cm), ECG-triggered non-shuttle mode (fixed table position) with stationary tube rotation (1 acquisition every 0.5-1 s, 5.8 cm), and non-ECG-triggered continuous mode (1 acquisition every 0.06 s, 5.8 cm). Stenosis was created in the circumflex artery, inducing different fractional flow reserve values. Volume perfusion CT Myocardium software was used to analyze ECG-triggered scans. For the non-ECG triggered scans MASS research version was used combined with an in-house Matlab script. MBF (mL/g/min) was calculated for non-ischemic segments. True MBF was calculated using input flow and heart weight. Significant differences in MBF between shuttle, non-shuttle and continuous mode were found, with median MBF of 0.87 [interquartile range 0.72-1.00], 1.20 (1.07-1.30) and 1.65 (1.40-1.88), respectively. The median MBF in shuttle mode was 56% lower than the true MBF. In non-shuttle and continuous mode, the underestimation was 41% and 18%. Limited temporal sampling rate in standard dynamic CTMPI techniques contributes to substantial underestimation of true MBF.Entities:
Keywords: Cardiac imaging techniques; Computed tomography; Perfusion imaging
Mesh:
Year: 2018 PMID: 30284642 PMCID: PMC6454077 DOI: 10.1007/s10554-018-1451-9
Source DB: PubMed Journal: Int J Cardiovasc Imaging ISSN: 1569-5794 Impact factor: 2.357
Fig. 1The arterial input functions (AIF) of shuttle (blue), non-shuttle (red) and continuous (green) perfusion mode with correlating CT images (Light grey square)
Fig. 2The median myocardial blood flow (MBF) in mL/min/g and interquartile range (IQR) for non-ischemic segments per heart for shuttle, non-shuttle and continuous CT perfusion scan mode
Fig. 3The median myocardial blood flow (MBF) in mL/min/g and interquartile range (IQR) for the different stenosis grades per scan mode for non-ischemic segments and the input flow in mL/min for heart 1 (a), heart 2 (b) and heart 3 (c)