| Literature DB >> 30283201 |
David L Driskell1, J Barton Gillum2, Jonathan D Monti2, Aaron Cronin2.
Abstract
OBJECTIVE: The study's primary objective was to determine army medics' accuracy performing bedside ultrasound (US) to detect radiolucent foreign bodies (FBs) in a soft-tissue hand model. Secondary objectives included the assessment of US stand-off pad effects on soft-tissue FB detection rates and assess established FB detectable lower limit size of 2 mm.Entities:
Keywords: Army medics; chicken model; foreign bodies; stand-off pad; ultrasound
Year: 2018 PMID: 30283201 PMCID: PMC6159331 DOI: 10.4103/JMU.JMU_12_18
Source DB: PubMed Journal: J Med Ultrasound ISSN: 0929-6441
Figure 1Thenar Eminence Model picture
Test performance by FB size (1-3 mm) without stand-off pad (n=280) and with stand-off pad (n=260)
| Modality | Size (mm) | Totals scans | FB present | Found | Missed | 95% CI | |
|---|---|---|---|---|---|---|---|
| Linear probe alone | 1-3 | 280 | 168 | 123 | 45 | 66.0-79.3 | 0.003 |
| Linear probe + stand-off pad | 260 | 156 | 90 | 66 | 49.8-65.1 | ||
| Linear probe alone | 1 | 168 | 56 | 36 | 20 | 37.0-59.2 | 0.1203 |
| Linear probe + stand-off pad | 160 | 52 | 25 | 27 | 51.1-75.5 | ||
| Linear probe alone | 2 | 168 | 56 | 46 | 10 | 52.0-73.5 | 0.032 |
| Linear probe + stand-off pad | 160 | 52 | 33 | 19 | 69.9-90.2 | ||
| Linear probe alone | 3 | 168 | 56 | 41 | 15 | 47.9-73.5 | 0.221 |
| Linear probe + stand-off pad | 160 | 52 | 32 | 20 | 60.3-83.1 | ||
| Linear probe alone | No FB | 280 | 112 | 25 | 87 | 15.5-30.9 | 0.025 |
| Linear probe + stand-off pad | 260 | 104 | 38 | 66 | 27.9-46.1 | ||
Fisher exact χ2. FP: False positive, TN: True negative, FB: Foreign body, CI: Confidence interval
Test performance characteristics for ultrasound without (n=280) and with stand-off pad (n=260)
| Modality | Percentage (95% CI) | |||
|---|---|---|---|---|
| Sensitivity | Specificity | PPV | NPV | |
| Linear probe alone | 73 (65.7-79.6) | 78 (68.6-84.7) | 83 (75.8-88.5) | 65 (57.0-73.7) |
| Linear probe + stand-off pad | 58 (49.5-65.5) | 67 (57.3-75.9) | 72 (63.7-80.0) | 51 (42.7-60.0) |
| Linear probe alone ≥2 mm | 78 (68.6-84.7) | 78 (68.6-84.7) | 78 (68.6-84.7) | 78 (68.6-84.7) |
| Linear probe + stand-off pad ≥2 mm | 62 (52.4-71.6) | 67 (57.3-75.9) | 65 (55.3-47.7) | 64 (54.4-73.0) |
CI: Confidence interval, PPV: Positive-predictive value, NPV: Negative predictive value
Comparison of health care providers
| Provider | Size (mm) | Accuracy (%) | Sensitivity | Specificity | PPV | NPV |
|---|---|---|---|---|---|---|
| Medic | 2-3 | 78 | 78 | 78 | 78 | 78 |
| Emergency medicine residents | 85 | 82 | 83 | 85 | ||
| Emergency physician | 15 | 80 | 70 | 90 | 88 | 75 |
| 15 | 77 | 69 | 84 | 82 | 73 | |
| 15 | 84 | 82 | 87 | 86 | 83 | |
| Radiologist | 15 | 83 | 83 | 83 | 83 | 83 |
| Nurse practitioner | 15 | 78 | 50 | 83 | 44 | |
| Emergency medicine physicians | 15 | 83 | 75 | 91 | 60 |
PPV: Positive predictive value, NPV: Negative-predictive value, FB: Foreign body