| Literature DB >> 30279909 |
Shinya Sugiura1, Koji Matsuoka1, Akiyoshi Ikami1, Hirokazu Shioji1, Akihiro Takasaki1, Takafumi Koji1, Takashi Tanigawa1, Masaaki Ito2.
Abstract
Cardiac pacemakers using rate response technologies represent the effectiveness of increasing the heart rate and cardiac output during exercise. Minute ventilation (MV) sensors are popular and estimate rates using transthoracic impedance by emitting very low amperage short electrical current pulses between the pacemaker lead tip and pulse generator. We present a case of an incomplete pacemaker lead fracture developing inappropriate pacing inhibition due to oversensing caused by the electrical current emitted by the MV sensor. A permanent pacemaker replacement was performed, resulting in no further abnormal findings such as inappropriate pacing inhibition. <Learning objective: The minute ventilation (MV) sensor is estimated using the transthoracic impedance by emitting a very low amperage and short pulses of electrical current between the pacemaker lead tip and pulse generator. Therefore, if a fractured lead occurs, the voltage from the current pulses emitted for the MV sensor might be high enough to be oversensed as ventricular waves. We recommend that it is necessary to keep in mind the possibility of a fractured lead if oversensing of the thoracic impedance measurement current is observed.>.Entities:
Keywords: Cardiac pacemaker; Incomplete lead fracture; Lead impedance; Minute ventilation sensor; Oversensing
Year: 2018 PMID: 30279909 PMCID: PMC6149620 DOI: 10.1016/j.jccase.2018.04.005
Source DB: PubMed Journal: J Cardiol Cases ISSN: 1878-5409