Literature DB >> 30279909

An incomplete pacemaker lead fracture causing inappropriate pacing inhibition due to oversensing of the minute ventilation sensor pulses.

Shinya Sugiura1, Koji Matsuoka1, Akiyoshi Ikami1, Hirokazu Shioji1, Akihiro Takasaki1, Takafumi Koji1, Takashi Tanigawa1, Masaaki Ito2.   

Abstract

Cardiac pacemakers using rate response technologies represent the effectiveness of increasing the heart rate and cardiac output during exercise. Minute ventilation (MV) sensors are popular and estimate rates using transthoracic impedance by emitting very low amperage short electrical current pulses between the pacemaker lead tip and pulse generator. We present a case of an incomplete pacemaker lead fracture developing inappropriate pacing inhibition due to oversensing caused by the electrical current emitted by the MV sensor. A permanent pacemaker replacement was performed, resulting in no further abnormal findings such as inappropriate pacing inhibition. <Learning objective: The minute ventilation (MV) sensor is estimated using the transthoracic impedance by emitting a very low amperage and short pulses of electrical current between the pacemaker lead tip and pulse generator. Therefore, if a fractured lead occurs, the voltage from the current pulses emitted for the MV sensor might be high enough to be oversensed as ventricular waves. We recommend that it is necessary to keep in mind the possibility of a fractured lead if oversensing of the thoracic impedance measurement current is observed.>.

Entities:  

Keywords:  Cardiac pacemaker; Incomplete lead fracture; Lead impedance; Minute ventilation sensor; Oversensing

Year:  2018        PMID: 30279909      PMCID: PMC6149620          DOI: 10.1016/j.jccase.2018.04.005

Source DB:  PubMed          Journal:  J Cardiol Cases        ISSN: 1878-5409


  3 in total

1.  Rate responsive cardiac pacing using a minute ventilation sensor.

Authors:  A M Abrahamsen; S Barvik; T Aarsland; K Dickstein
Journal:  Pacing Clin Electrophysiol       Date:  1993-08       Impact factor: 1.976

2.  Unexpected and unmanageable malfunctions of current co-radial pacemaker leads.

Authors:  Masayuki Ishimura; Marehiko Ueda; Kazuo Miyazawa; Takatsugu Kajiyama; Naotaka Hashiguchi; Masahiro Nakano; Yusuke Kondo; Tomonori Kanaeda; Yasunori Hiranuma; Yoshio Kobayashi
Journal:  Europace       Date:  2014-08-08       Impact factor: 5.214

3.  Initial clinical experience with a minute ventilation sensing rate modulated pacemaker: improvements in exercise capacity and symptomatology.

Authors:  C P Lau; A Antoniou; D E Ward; A J Camm
Journal:  Pacing Clin Electrophysiol       Date:  1988-11       Impact factor: 1.976

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.