Enhancement of capnogram waveform in the presence of chest compression artefact during cardiopulmonary resuscitation.

Sofía Ruiz de Gauna1, Mikel Leturiondo2, J Julio Gutiérrez2, Jesus M Ruiz2, Digna M González-Otero2, James K Russell3, Mohamud Daya3.   

Abstract

BACKGROUND: Current resuscitation guidelines emphasize the use of waveform capnography to help guide rescuers during cardiopulmonary resuscitation (CPR). However, chest compressions often cause oscillations in the capnogram, impeding its reliable interpretation, either visual or automated. The aim of the study was to design an algorithm to enhance waveform capnography by suppressing the chest compression artefact.
METHODS: Monitor-defibrillator recordings from 202 patients in out-of-hospital cardiac arrest were analysed. Capnograms were classified according to the morphology of the artefact. Ventilations were annotated using the transthoracic impedance signal acquired through defibrillation pads. The suppression algorithm is designed to operate in real-time, locating distorted intervals and restoring the envelope of the capnogram. We evaluated the improvement in automated ventilation detection, estimation of ventilation rate, and detection of excessive ventilation rates (over-ventilation) using the capnograms before and after artefact suppression.
RESULTS: A total of 44 267 ventilations were annotated. After artefact suppression, sensitivity (Se) and positive predictive value (PPV) of the ventilation detector increased from 91.9/89.5% to 98.0/97.3% in the distorted episodes (83/202). Improvement was most noticeable for high-amplitude artefact, for which Se/PPV raised from 77.6/73.5% to 97.1/96.1%. Estimation of ventilation rate and detection of over-ventilation also upgraded. The suppression algorithm had minimal impact in non-distorted data.
CONCLUSION: Ventilation detection based on waveform capnography improved after chest compression artefact suppression. Moreover, the algorithm enhances the capnogram tracing, potentially improving its clinical interpretation during CPR. Prospective research in clinical settings is needed to understand the feasibility and utility of the method.
Copyright © 2018 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Advanced life support; Cardiopulmonary resuscitation; Chest compression artefact; Chest compressions; Ventilation; Waveform capnography

Mesh:

Year:  2018        PMID: 30278204     DOI: 10.1016/j.resuscitation.2018.09.024

Source DB:  PubMed          Journal:  Resuscitation        ISSN: 0300-9572            Impact factor:   5.262


  2 in total

1.  Laryngeal mask versus facemask in the respiratory management during catheter ablation.

Authors:  Takashi Koyama; Masanori Kobayashi; Tomohide Ichikawa; Yasushi Wakabayashi; Daiki Toma; Hidetoshi Abe
Journal:  BMC Anesthesiol       Date:  2020-01-07       Impact factor: 2.217

2.  Incremental effect of non-invasive oscillating device on chest physiotherapy in critically ill children: a cross-over randomised trial.

Authors:  Atsushi Kawaguchi; Gabrielle Bernier; Andy Adler; Guillaume Emeriaud; Philippe A Jouvet
Journal:  BMJ Open       Date:  2020-10-05       Impact factor: 2.692

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.