Physical Activities That Cause High Friction Moments at the Cup in Hip Implants.

Georg Bergmann1, Alwina Bender1, Jörn Dymke1, Georg N Duda1, Philipp Damm1.   

Abstract

BACKGROUND: High friction moments in hip implants contribute to the aseptic loosening of cementless cups, of which there are approximately 100,000 cases per year; sustained joint loading may cause such high moments. The most "critical" physical activities associated with sustained joint loading were identified in this study.
METHODS: Friction moments in the cup were telemetrically measured about 33,000 times in the endoprostheses of 9 subjects during >1,400 different activities. The highest moments were compared with the cup's fixation stability limit of approximately 4 Nm.
RESULTS: A total of 124 different activities caused friction moments meeting or exceeding the critical limit, with the highest value of 11.5 Nm. Most involved sustained high contact forces before or during the activity. The highest peak moments (6.3 to 11.5 Nm) occurred when moving the contralateral leg during 1-legged stance, during breaststroke swimming, muscle stretching, 2-legged stance with muscle contraction, and during static 1-legged stance. The median moments were highest (3.4 to 3.9 Nm) for unstable 1-legged stance, whole-body vibration training, 2-legged stance with an unexpected push at the upper body, 1-legged stance while exercising the contralateral leg, and running after 2-legged stance.
CONCLUSIONS: Frequent unloading plus simultaneous movement of the joint are required to maintain good joint lubrication and keep the friction moments low. Frequent, sustained high loads before or during an activity may cause or contribute to aseptic cup loosening. During the first months after hip arthroplasty, such activities should be avoided or reduced as much as possible. This especially applies during postoperative physiotherapy. Whether these guidelines also apply for subjects with knee implants or arthrotic hip or knee joints requires additional investigation. CLINICAL RELEVANCE: The risk of aseptic cup loosening may be reduced by avoiding sustained loading of hip implants without periodic joint movement.

Entities:  

Mesh:

Year:  2018        PMID: 30277993     DOI: 10.2106/JBJS.17.01298

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  7 in total

1.  Analysis of hip joint loading during walking with different shoe types using instrumented total hip prostheses.

Authors:  Y Palmowski; S Popović; D Kosack; P Damm
Journal:  Sci Rep       Date:  2021-05-12       Impact factor: 4.379

2.  Investigation of Regression Methods for Reduction of Errors Caused by Bending of FSR-Based Pressure Sensing Systems Used for Prosthetic Applications.

Authors:  Chakaveh Ahmadizadeh; Carlo Menon
Journal:  Sensors (Basel)       Date:  2019-12-13       Impact factor: 3.576

3.  Recommendations for return to sports after total hip arthroplasty are becoming less restrictive as implants improve.

Authors:  T Vu-Han; S Hardt; R Ascherl; C Gwinner; C Perka
Journal:  Arch Orthop Trauma Surg       Date:  2020-12-01       Impact factor: 3.067

4.  Finite element analysis of femoral neck strains during stair ascent and descent.

Authors:  Jason C Gillette; Timothy R Derrick; Chen Deng
Journal:  Sci Rep       Date:  2021-04-28       Impact factor: 4.379

5.  In vivo analysis of hip joint loading on Nordic walking novices.

Authors:  Yannick Palmowski; Srdan Popovic; Simone G Schuster; Sebastian Hardt; Philipp Damm
Journal:  J Orthop Surg Res       Date:  2021-10-14       Impact factor: 2.359

6.  Overstretching Expectations May Endanger the Success of the "Millennium Surgery".

Authors:  Alwina Bender; Philipp Damm; Hagen Hommel; Georg N Duda
Journal:  Front Bioeng Biotechnol       Date:  2022-02-14

7.  Surgical cup placement affects the heating up of total joint hip replacements.

Authors:  Philipp Damm; Alwina Bender; Vivian Waldheim; Tobias Winkler; Georg N Duda
Journal:  Sci Rep       Date:  2021-08-04       Impact factor: 4.379

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.