Literature DB >> 30274947

Cumulative Long-Term Incidence of Postoperative Periprosthetic Femoral Fractures Using an Uncemented Tapered Titanium Hip Stem: 26- to 32-Year Results.

David S Peitgen1, Moritz M Innmann1, Christian Merle1, Tobias Gotterbarm2, Babak Moradi1, Marcus R Streit1.   

Abstract

BACKGROUND: Although falling is becoming a major problem in the elderly, little is known about the long-term incidence of postoperative periprosthetic femoral fractures (PFF) after total hip arthroplasty (THA) using uncemented stems with a minimum follow-up of 26 years postoperatively.
METHODS: In a consecutive series, the cumulative incidence of PFF after uncemented THA using a straight, collarless, tapered titanium stem was retrospectively reviewed in 354 hips (326 patients). After a mean follow-up of 28 (26-32) years postoperatively, 179 hips had died and 5 hips had been lost to follow-up. Kaplan-Meier survival analysis was used to estimate cumulative percentage probability of PFF.
RESULTS: At final follow-up, a total of 27 fractures in 27 patients had occurred. In 15 hips, the stem had to be revised, and in 10 the fracture was treated by open reduction and internal fixation. Two patients declined further surgery. The cumulative percentage probability of PFF was 1.6% (95% confidence interval, 0.7 to 3.8) at 10 years and 13.2% (95% confidence interval, 8.5 to 20.2) at 29 years after primary THA. There was no association between the occurrence of fracture and gender, age at primary THA, cup revision, or canal fill index.
CONCLUSION: Our findings indicate that PFF is the major mode of failure in the long term after uncemented THA using a tapered titanium stem. More research is needed to evaluate the reasons and risk factors for PFF after THA, thus providing better prophylaxis for those at risk and to compare the long-term incidence using different fixation methods and implants. LEVEL OF EVIDENCE: IV.
Copyright © 2018 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  CLS Spotorno stem; cementless total hip arthroplasty; long-term follow-up; periprosthetic femoral fracture; uncemented total hip arthroplasty

Mesh:

Substances:

Year:  2018        PMID: 30274947     DOI: 10.1016/j.arth.2018.08.038

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  3 in total

1.  Periprosthetic proximal femoral fractures in cemented and uncemented stems according to Vancouver classification: observation of a new fracture pattern.

Authors:  James Karam; Paul Campbell; Shivang Desai; Michael Hunter
Journal:  J Orthop Surg Res       Date:  2020-03-10       Impact factor: 2.359

2.  Long-Term Results of Hip Arthroplasty Using Extensive Porous-Coated Stem-A Minimum Follow-Up of 15 Years.

Authors:  Myung Hoon Park; Yung Hun Youn; Joon Soon Kang; Kyoung Ho Moon
Journal:  Geriatr Orthop Surg Rehabil       Date:  2019-12-19

3.  Effects of porous tantalum on periprosthetic bone remodeling around metaphyseal filling femoral stem: a multicenter, prospective, randomized controlled study.

Authors:  Goro Motomura; Naohiko Mashima; Hiroshi Imai; Akihiro Sudo; Masahiro Hasegawa; Harumoto Yamada; Mitsuhiro Morita; Naoto Mitsugi; Ryosuke Nakanishi; Yasuharu Nakashima
Journal:  Sci Rep       Date:  2022-01-18       Impact factor: 4.379

  3 in total