Evaluation of a pneumatic surgical robot with dynamic force feedback.

Dimitrios Karponis1,2, Yokota Koya3, Ryoken Miyazaki3, Takahiro Kanno3, Kenji Kawashima3.   

Abstract

Robot-assisted surgery is limited by the lack of haptic feedback and increased operating times. Force scaling adjusts feedback transmitted to the operator through the use of scaling factors. Herein, we investigate how force scaling affects forces exerted in robotic surgery during simple and complex tasks, using a pneumatic surgical robot, IBIS VI. Secondary objectives were to test the effects of force scaling on operating time, depth of needle insertion and user satisfaction. Two novice males performed simple (modified block transfer) and complex (needle insertion) tasks under four scaling factors: 0.0, 0.5, 1.0 and 2.0. Single-blind experiments were repeated five times, with alternating scaling factors in random order. Increasing the scaling factor from 0.0 to 2.0 reduces forces in block transfer (p = 0.04). All feedback conditions reduce forces in needle insertion compared to baseline (0.5: p < 0.001, 1.0: p = 0.001, 2.0: p = 0.001). Time to complete block transfer is shorter for scaling factor 0.5 (p = 0.02), but not for 1.0 (p = 0.05) or 2.0 (p = 0.48), compared to baseline. Depth of needle insertion decreases consistently with incremental scaling factors (p < 0.001). Further reductions are observed upon augmenting feedback (0.5-2.0: p = 0.02). User satisfaction in block transfer is highest for intermediate scaling factors (0.0-1.0: p = 0.01), but no change is observed in needle insertion (p = 0.99). Increments in scaling factor reduce forces exerted, particularly in tasks requiring precision. Depth of needle insertion follows a similar pattern, but operating time and user satisfaction are improved by intermediate scaling factors. In summary, dynamic adjustment of force feedback can improve operative outcomes and advance surgical automation.

Keywords:  Block transfer; Force feedback; Force scaling; Needle insertion; Pneumatic surgical robot; Robot-assisted surgery

Mesh:

Year:  2018        PMID: 30255361     DOI: 10.1007/s11701-018-0878-2

Source DB:  PubMed          Journal:  J Robot Surg        ISSN: 1863-2483


  2 in total

1.  Three-dimensional posture estimation of robot forceps using endoscope with convolutional neural network.

Authors:  Takuto Mikada; Takahiro Kanno; Toshihiro Kawase; Tetsuro Miyazaki; Kenji Kawashima
Journal:  Int J Med Robot       Date:  2020-01-08       Impact factor: 2.547

2.  Pneumatically driven surgical forceps displaying a magnified grasping torque.

Authors:  Takuya Iwai; Takahiro Kanno; Tetsuro Miyazaki; Daisuke Haraguchi; Kenji Kawashima
Journal:  Int J Med Robot       Date:  2020-01-05       Impact factor: 2.547

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.