Radiation Dose Reduction With a Low-Tube Voltage Technique for Pediatric Chest Computed Tomographic Angiography Based on the Contrast-to-Noise Ratio Index.

Takanori Masuda1, Yoshinori Funama2, Takeshi Nakaura3, Masahiro Tahara4, Yukari Yamashita5, Masao Kiguchi6, Naoyuki Imada5, Tomoyasu Sato7, Kazuo Awai6.   

Abstract

INTRODUCTION: The aim of this study was to evaluate the radiation dose and image quality at low tube-voltage pediatric chest computed tomographic angiography (CTA) that applies the same contrast-to-noise ratio (CNR) index as the standard tube voltage technique.
MATERIALS AND METHODS: Contrast-enhanced chest CTA scans of 100 infants were acquired on a 64-row multidetector computed tomography (MDCT) scanner. In the retrospective study, we evaluated 50 images acquired at 120 kVp; the image noise level was set at 25 Hounsfield units. In the prospective study, we used an 80-kVp protocol; the image noise level was 40 Hounsfield units because the iodine contrast was 1.6 times higher than on 120-kVp scans; the CNR was as in the 120-kVp protocol. We compared the CT number, image noise, CT dose index volume (CTDIvol), and the dose-length product on scans acquired with the 2 protocols. A diagnostic radiologist and a pediatric cardiologist visually evaluated all CTA images.
RESULTS: The mean CTDIvol and the mean dose-length product were 0.5 mGy and 7.8 mGy-cm for 80- and 1.2 mGy and 20.8 mGy-cm for 120-kVp scans, respectively (P < .001). The mean CTDIvol was 42% lower at 80 kVp than at 120 kVp, and there was no significant difference in the visual scores assigned to the CTA images (P = .28).
CONCLUSIONS: With the CNR index being the same at 80-kVp and 120-kVp imaging, the radiation dose delivered to infants subjected to chest CTA can be reduced without degradation of the image quality.
Copyright © 2018 Canadian Association of Radiologists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Contrast-to-noise ratio; Low-tube-voltage; Pediatric computed tomographic angiography; Radiation dose reduction

Mesh:

Year:  2018        PMID: 30249409     DOI: 10.1016/j.carj.2018.05.004

Source DB:  PubMed          Journal:  Can Assoc Radiol J        ISSN: 0846-5371            Impact factor:   2.248


  1 in total

1.  Variation in tube voltage for pediatric neck 64VCT: Effect on radiation dose and image quality.

Authors:  Li-Guo Chen; Ping-An Wu; Hsing-Yang Tu; Ming-Huei Sheu; Li-Chuan Huang
Journal:  PLoS One       Date:  2021-11-12       Impact factor: 3.240

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.