| Literature DB >> 30234374 |
Susan LaRue1, Jane Springer1, Michael Noderer2, James Meehan2, Carol H Wysham3.
Abstract
BACKGROUND: Ease of injection is important to patients. An autoinjector was developed to deliver exenatide, a glucagon-like peptide-1 receptor agonist for type 2 diabetes mellitus. For autoinjection, 0.06-mm exenatide-containing microspheres are suspended in medium-chain triglycerides. Herein, we report design verification and usability testing of the autoinjector for exenatide once-weekly suspension (QWS) delivery.Entities:
Keywords: exenatide; exenatide once-weekly suspension autoinjector; type 2 diabetes mellitus; usability; validation study
Mesh:
Substances:
Year: 2018 PMID: 30234374 PMCID: PMC6399794 DOI: 10.1177/1932296818798376
Source DB: PubMed Journal: J Diabetes Sci Technol ISSN: 1932-2968
Number of Discrete Use Steps and Comparison of Major Steps Required for Administration of Exenatide Once Weekly as a Single-Dose Injection Tray, Dual-Chamber Pen, or Suspension for Autoinjection.
| Single-dose injection tray | Dual-chamber pen | Suspension for autoinjection | |
|---|---|---|---|
| Number of discrete use steps[ | 35 | 16 | 13 |
| Major steps[ | • Prepare vial | • Assemble device (attach needle to pen) | • Mix |
Summarized and described based on the instructions for use for each device.
Key Design Goals of Exenatide Once-Weekly Suspension Autoinjector.
| Design goal | Strategies for and evidence of achievement |
|---|---|
| Simpler administration relative to the dual-chamber pen | • Device requires fewer overall steps to administer |
| Comfortable to use (size and shape for convenient transporting and handling, acceptable torque/force requirements) | • Device has a square body with rounded edges that is ergonomic |
| Easy to use | • 78% of tested participants successfully completed an injection within the summative study |
| Cartridge contents visible to user | • Device includes a large viewing window |
| Fixed, single dose | • Device contains a single dose |
| Facilitation of mixing, purging of air from the cartridge, and complete dose delivery | • Large viewing window enables users to ensure suspension is properly mixed |
| Safe and sterile containment of needle | • Needle is hidden by a needle shield and kept sterile until activation at time of injection |
Figure 1.Exenatide once-weekly suspension autoinjector device before and after use.
Performance on Critical Tasks (N = 104).
| Task | Success,[ | Success with difficulty,[ | Use error,[ | Success with or without difficulty,[ |
|---|---|---|---|---|
| Differentiation scenarios | ||||
| Visually distinguish and select correct device from among other injection devices (loose packaging)[ | 79 | 3[ | 7[ | 92 |
| Visually distinguish and select correct device from among other injection devices | 100 | 1 | 3[ | 97 |
| Injection scenarios | ||||
| Select the appropriate injection site | 103 | 1 | 0 | 100 |
| Mix the medication by shaking | 102 | 1 | 1 | 99 |
| Check consistency of mixture to determine if well mixed; shake more if not mixed | 103 | 0 | 1 | 99 |
| Remove needle cap | 65 | 31 | 8 | 92 |
| Complete injection of well-mixed product[ | 89 | 0 | 13[ | 87 |
| Inject the dose | 96 | 3 | 5[ | 95 |
| Hold to skin for complete injection | 99 | 0 | 5 | 95 |
| Post-scenario questions about storage of autoinjector | ||||
| Properly store autoinjector | 104[ | 0 | 0 | 100 |
| Use of device label or carton to answer comprehension questions | ||||
| Finding/understanding expiration date | 100[ | 4[ | 0 | 100 |
| Frequency of injection | 104[ | 0 | 0 | 100 |
| Use of the instructions for use to answer comprehension questions | ||||
| Do not use expired autoinjectors | 96[ | 8 | 0 | 100 |
| Inject on the skin | 104[ | 0 | 0 | 100 |
| Confirm well-mixed medication through inspection of autoinjector window | 103[ | 0 | 0 | 100 |
Participants were asked to perform only one injection attempt. Ratings are for this single, initial attempt.
Retail pharmacists did not complete this scenario because pharmacies stock full packaging carton; therefore, the denominator is 89 for this task.
Errors were due to study artifact; artifacts were defined as errors that were not attributed to the design interface but attributed to the study itself (ie, unable to simulate real-world conditions completely).
Two mixes could not be observed by the moderator because of premature dispensing of medication; therefore, the denominator is 102 for this task.
One use error was due to study artifact.
Three use errors were due to study artifact; two participants recapped but continued with the needle shield locked.
Correct response.
One participant was not asked this question.
Design Verification Testing Under Standard Atmospheric Conditions[a] (N = 60).
| System characteristic | Specification range | Minimum | Maximum | Mean ± SD | Result |
|---|---|---|---|---|---|
| Dose accuracy, mL | 0.8075-0.8925 | 0.8319 | 0.8482 | 0.8402 ± 0.0039 | Pass |
| Torque to unlock, Nm | 0.023-0.250 | 0.109 | 0.182 | 0.152 ± 0.018 | Pass |
| Torque to uncap, Nm | 0.023-0.250 | 0.117 | 0.199 | 0.145 ± 0.016 | Pass |
| Force to depress needle shield and start injection, N | 5.00-15.00 | 9.77 | 11.32 | 10.65 ± 0.37 | Pass |
Standard atmospheric conditions: temperature, 18-28°C; relative humidity, 25-75%.
SD, standard deviation.
Participant Demographics.
| Demographics, n (%) | Lay users (n = 73) | Health care professionals (n = 16) | Retail pharmacists (n = 15) |
|---|---|---|---|
| Sex | |||
| Female | 34 (47) | 7 (44) | 6 (40) |
| Male | 39 (53) | 9 (56) | 9 (60) |
| Age, years | |||
| 18-54 | 18 (25) | NA | NA |
| 55-69 | 46 (63) | NA | NA |
| 70-75 | 9 (12) | NA | NA |
| Race/ethnicity | |||
| Caucasian | 41 (56) | 5 (31) | 7 (47) |
| Black | 11 (15)[ | 2 (13) | 1 (7) |
| Hispanic | 13 (18)[ | 5 (31) | 0 (0) |
| Asian | 3 (4) | 2 (13) | 1 (7) |
| Middle Eastern | 1 (1) | 0 (0) | 5 (33) |
| Other/unknown | 5 (7) | 2 (13) | 1 (7) |
| Education | |||
| High school graduate | 2 (3) | NA | NA |
| Some college | 21 (29) | NA | NA |
| College graduate | 36 (49) | NA | NA |
| Postgraduate | 9 (12) | NA | NA |
| Unknown | 5 (7) | NA | NA |
| Right-handed | 62 (85) | 15 (94) | 15 (100) |
| Colorblind | 2 (3) | 0 (0) | 0 (0) |
| Wear glasses | 57 (78) | 12 (75) | 3 (20) |
| Patient/caregiver status | |||
| Patients | 61 (84) | NA | NA |
| Caregivers | 12 (16) | NA | NA |
| Injection naïve[ | 32 (44) | NA | NA |
| Administer injections[ | NA | 12 (75) | 4 (27) |
One lay user was recorded as black race and Hispanic ethnicity.
Lay users were categorized as injection experienced or naïve. Health care professionals and pharmacists were categorized by whether they administered injections.
NA, not applicable.
Figure 2.Participant success on key (A) critical and (B) essential tasks with and without difficulty.
aSuccessful injection of a complete dose required success on the three criteria shown: completing injection of a well-mixed product, injecting the dose, and holding the device to the skin for complete injection.
Figure 3.Participant ease-of-use ratings by task (N = 104). Participants were only asked about the tasks they completed. No participants rated any task as “very difficult.”
Unique Steps in the Instructions for Use of Exenatide Once-Weekly Suspension Autoinjector.
| Instruction | Rationale |
|---|---|
| Store the autoinjector flat | The medication will be easier to mix when the autoinjector has been stored flat because the microspheres will settle over a larger surface area |
| Remove the autoinjector from the refrigerator at least 15 minutes before use | The medication is easier to mix well at warmer temperatures |
| Shake the autoinjector hard for at least 15 seconds | Shaking is required to mix the suspension well |
| Shake the autoinjector until the medicine is well-mixed and cloudy | A cloudy appearance indicates thorough mixing as the microspheres are in suspension |
| When unlocking the autoinjector and unscrewing the orange cap, the needle and orange cap should be pointing up | The needle-up position allows the microspheres to stay within the syringe |
| Inject the medicine as soon as the mixing is complete; do not unlock the autoinjector until just before injection | Injecting the medication as soon as the mixing is complete ensures the microspheres remain in suspension |
| After pushing the autoinjector against the skin, keep holding it there for 15 seconds | Holding the autoinjector in place for 15 seconds allows sufficient time to deliver the required volume |