Literature DB >> 30221222

Self-Calibration Phenomenon for Near-Infrared Clinical Measurements: Theory, Simulation, and Experiments.

Idit Feder1, Hamootal Duadi1, Ruchira Chakraborty1, Dror Fixler1.   

Abstract

An irradiated turbid medium scatters the light in accordance to its optical properties. Near-infrared (NIR) clinical methods, which are based on spectral-dependent absorption, suffer from an inherent error due to spectral-dependent scattering. We present here a unique spatial point, that is, iso-pathlength (IPL) point, on the surface of a tissue at which the intensity of re-emitted light remains constant. This scattering-indifferent point depends solely on the medium geometry. On the basis of this natural phenomenon, we suggest a novel optical method for self-calibrated clinical measurements. We found that the IPL point exists in both cylindrical and semi-infinite tissue geometries (Supporting Information, Video file). Finally, in vivo human finger and mice measurements are used to validate the crossing point between the intensity profiles of two wavelengths. Hence, measurements at the IPL point yield an accurate absorption assessment while eliminating the scattering dependence. This finding can be useful for oxygen saturation determination, NIR spectroscopy, photoplethysmography measurements, and a wide range of optical sensing methods for physiological aims.

Entities:  

Year:  2018        PMID: 30221222      PMCID: PMC6130783          DOI: 10.1021/acsomega.8b00018

Source DB:  PubMed          Journal:  ACS Omega        ISSN: 2470-1343


Introduction

Diagnosis of a physiological state by extracting optical parameters of tissues, as implemented in near-infrared (NIR) spectroscopy, optical coherence tomography, fluorescence, etc.,[1−5] is widespread. However, human tissue, which is a turbid medium, poses a challenge for the different optical methods. Light reflectance from a tissue has spectral-dependent absorption and scattering.[6] Hence, evaluating absorption-dependent physiological parameters requires calibration of the scattering to obtain accurate information. Methods for extracting optical properties from the reflected light are based on either the amplitude in multiple sources or multiple point measurements. Most methods are complex and based on modulated sources with phase, time, and frequency analyses.[7] Pulse oximetry is one of the most common point-of-care tests because it is noninvasive, characterized by high speed, and can be interpreted even without a physician. It uses two light sources for extracting blood saturation, pressure, and perfusion on the basis of ratiometric absorption measurements. The change in optical properties due to wavelength produces a difference in the optical path. Hence, at different wavelengths, a position on the surface of the medium has an intensity that is influenced by a different optical path. The full scattering profile (FSP), which is the angular intensity distribution,[8,9] is a new point of view for extracting optical properties of cylindrical tissues using a simple system containing a continuous wave source and a photodetector (PD). The FSP of cylindrical tissues is relevant to biomedical measurement of fingers, earlobes, or pinched tissues. Using the FSP, the iso-pathlength (IPL) point phenomenon is revealed; an IPL point is a point on the surface of a cylindrical medium where the light intensity remains constant and does not depend on the reduced scattering coefficient of the medium but on the spatial structure. This phenomenon serves as a self-calibration point because the intensity there is influenced by absorption regardless of the scattering and thus improves the accuracy of the optical property extraction. Moreover, we presented a linear dependency between the radius of the tissue and the point’s location. Most existing instruments do not deal with the FSP but rather focus only on the reflection or transmission of light. The novel finding of the IPL point, to the best of our knowledge, has not been reported before this research. However, we have found supporting reports on reflection from a semi-infinite medium in diffusion reflection (DR)[10,11] and reflection phase measurements.[12] Mourant[10] reported on a source–detector separation that does not depend on scattering properties at the range close to the light source. Mourant presented a Monte Carlo (MC) simulation and phantom experiments of reflected light from a semi-infinite medium. Moreover, Calabro showed the influence of the phase function on generalized diffuse reflectance models[13] from a semi-infinite medium and presented a singular distance from the source, below the diffusion regime, which is insensitive to the phase function. Savo et al.[14] recently presented a research that supports our work. They reported that the light mean path length is invariant with respect to the microstructure of the medium. The work was based on the random walk theory, MC simulations, and measurements of scattered light from glass tubes with scattering solutions using an autocorrelator for extracting the mean path length. We believe that this natural phenomenon of a point, which is indifferent to changes in the scattering property, depends solely on the tissue geometry. Knowledge of the position of this point in each geometry will allow an accurate, self-calibrated detection of absorption-dependent physiological parameters. Note that this phenomenon only occurs when multiple scattering exists.[6] We show this phenomenon from a semi-infinite medium theoretically, by simulations, and experimentally. Diffusion theory is supported by MC simulation results as well as the experimental results with phantoms. We consider two geometries, a cylindrical model and a semi-infinite model, and present a point that is not dependent on the scattering coefficient. Both geometries are validated by in vivo measurements, a human finger for the cylindrical model and a mouse neck for the semi-infinite model. The finding of the IPL point can be useful for oxygen saturation determination, NIR spectroscopy, photoplethysmography (PPG) measurements, and a wide range of optical sensing methods for physiological aims in a more accurate way. The FSP utilizes the entire distribution profile rather than just parts of it by the complete characterization of the material and its geometry, thereby improving the existing systems and creating new methods.

Diffusion Theory Simulation

The propagation of light through a tissue is mostly described by the radiative transport equation (RTE). RTE is an approximation to Maxwell’s equations, based on the conservation of energy, describing the propagation of energyThe radiation (L(r, ŝ, t)) transported by photons in direction ŝ and to a distance r from the source at time t is represented by eq , where c is the speed of light. The extinction coefficient is μt = μa + μs′, where μs′ is the reduced scattering coefficient defined by μs′ = (1 – g)μs for anisotropy factor g and scattering coefficient μs. P(ŝ′, ŝ) is the phase function, which gives the probability for a photon traveling in direction ŝ to be scattered in direction ŝ′. Ω is a solid angle, and S(r, ŝ, t) represents the radiant source in the tissue. The radiation integrated over all directions is defined by the fluence rate [W/cm2][15]The most typical approach for solving the RTE analytically is the diffusion approximation.[16,17] Using the solution of diffusion theory at steady state, we can get the expression for fluence ψ(r)[J/cm2] in response to power Q located at r = 0 as followsEquation describes the steady-state diffusion of light from a point source of power P, where r is the position on the XY plane, D is the diffusion length, D = 1/(3(μa + μs′)), and δ is the optical penetration depth, . We determine Q to be 1 to examine the light transport independent of the source’s power. Because we are interested in developing light diffusion equations on the surface of a cylindrical tissue, we express the fluence on the edge of a circle and obtain the FSP at a cross section, with radius a, of the illuminated cylinder. We can define distance r of a point on the circumference from the light source (Figure ) by the law of cosines as followsOn the basis of eqs and 4, the steady-state diffusion of light on the surface of the cylinder is obtained. This fluence in accordance with μs′ for each θ at the circle surface is represented by the following equationThe change of the fluence per θ can predict θIPL, where .For the relevant μs′ range, the fluence depends much more on the cylindrical radius a than on μs′ (Figure ). Because we assume an angular accuracy of 5°, all of the μs′ yield an identical angle.
Figure 1

Geometric description of a point on the circumference in a circular cross section.

Figure 2

Dependency of θIPL on radius a and reduced scattering coefficient μs′. μs′ has a smaller influence on the θIPL value (less than 5°) than the radius a.

Geometric description of a point on the circumference in a circular cross section. Dependency of θIPL on radius a and reduced scattering coefficient μs′. μs′ has a smaller influence on the θIPL value (less than 5°) than the radius a.

Results

Full Scattering Profile of a Cylindrical Medium: Theory, Simulation, and Experiment Results

The FSPs of a cylindrical medium based on diffusion theory are presented in Figure . The reduced scattering coefficients are in the human skin range of 15−30 cm−1 and the radius 1 cm. The IPL point appears at 127°.
Figure 3

FSPs and the IPL point of cylindrical tissues according to diffusion theory (logarithmic scale).

FSPs and the IPL point of cylindrical tissues according to diffusion theory (logarithmic scale). The location of the IPL point on the surface of a cylinder is dependent on its radius. As the radius increases, the fixed-intensity point becomes closer to the light source. Because diffusion theory is relevant for distances larger than one or two transport mean free path (MFP’) from the light source,[15] where MFP′ = 1/(μa +μs′). The radii smaller than the MFP’ are not valid for this theory, and similar is the case with the large radii where the IPL point appears at high angles because diffusion theory is inaccurate near the source. At the relevant range of radii, which describes fingers and cylindrical tissues that the light can penetrate through, the IPL point depends on the tissue radius. Using our MC simulation, which is appropriate for the regions close to the source, we described the FSPs of cylindrical tissues with different reduced scattering coefficients of 14 to 26 cm–1 for 6.5 mm tissue radius, as shown in Figure a, and simulated these FSPs for various tissue diameters. The IPL point appeared in each set of tissues. Moreover, a linear dependency on tissue radius was determined, as shown in Figure b.
Figure 4

(a) FSPs and the IPL point from the MC simulation of a cylindrical finger tissue, 6.5mm radius. (b) Linear dependency of IPL point on the radius.

(a) FSPs and the IPL point from the MC simulation of a cylindrical finger tissue, 6.5mm radius. (b) Linear dependency of IPL point on the radius. We built a unique setup (Figure ) for noninvasive encircled measurements containing a continuous wave source and a portable photodetector (PD).
Figure 5

Illustration of the setup for measuring FSP using a portable photodetector (PD).

Illustration of the setup for measuring FSP using a portable photodetector (PD). Our experiment reveals an IPL point, in accordance with the simulation, that is independent of the optical properties. We used one continuous wave laser, a photodetector, and tissue-mimicking phantoms presenting different optical properties. The FSP and the IPL point that appeared at 115° are shown in Figure a. The linear dependence of the IPL point location on the radius of the cylindrical phantom is presented in Figure b.
Figure 6

Experimental results of an IPL point in a cylindrical medium. (a) FSPs of phantoms, with 0.6 cm radius and (b) linear dependency of IPL point on the radius.

Experimental results of an IPL point in a cylindrical medium. (a) FSPs of phantoms, with 0.6 cm radius and (b) linear dependency of IPL point on the radius. We measured a human finger in vivo in our system to validate the FSP behavior of cylindrical tissues. We used two wavelengths and obtained the FSPs of a human finger, which are presented in Figure . The effective radius of the finger is 0.6 cm, and the common point for these FSPs appears at 110°, whereas in the phantom results, at a similar effective radius, the IPL point is obtained at 115°. The human finger contains blood vessels, which add a non-negligible absorption. The absorption has a spectral dependence; therefore, the measurements are decreased for each wavelength separately and the common point is different from the phantoms’ IPL point.
Figure 7

FSP of a male index finger in 650 (crosses) and 780 nm (squares).

FSP of a male index finger in 650 (crosses) and 780 nm (squares).

Reflection from a Semi-Infinite Medium: Theory, Simulation, and Experiment Results

On the basis of diffusion theory, we calculated the reflected light intensity from a semi-infinite medium at distance r from the source. The reflection intensity from a semi-infinite medium with different reduced scattering coefficients, in the range of 15–30 cm–1, is shown in Figure a. A crossing point appears at 8 mm from the source. This location can also be called the IPL distance, rIPL, from the illumination point.
Figure 8

Reflection from a semi-infinite tissue: (a) reflection according to diffusion theory; a crossing point does not depend on scattering properties (blue crosses, orange squares, yellow diamonds, and purple circles represent μs′ at 15, 20, 25, and 30 cm–1, respectively). (b) MC simulation of a crossing point, which is indifferent to changes in the scattering property (orange solid, yellow dash-dotted, magenta dotted, and green dashed lines represent μs′ at 8, 10, 12, and 14 cm–1, respectively).

Reflection from a semi-infinite tissue: (a) reflection according to diffusion theory; a crossing point does not depend on scattering properties (blue crosses, orange squares, yellow diamonds, and purple circles represent μs′ at 15, 20, 25, and 30 cm–1, respectively). (b) MC simulation of a crossing point, which is indifferent to changes in the scattering property (orange solid, yellow dash-dotted, magenta dotted, and green dashed lines represent μs′ at 8, 10, 12, and 14 cm–1, respectively). In addition, a MC simulation for various reduced scattering coefficients, in the range of 8–14 cm–1, suggested that path lengths do not depend on the scattering properties of the medium at a distance of rIPL = 3.5 mm from the source, as presented in Figure b. These values represent the optical tissue coefficients, as we used in the experimental system. We conducted an experiment on DR measurements from a solid silicon-based phantom without absorption components.[18,19] An illustration of the measurement system is shown in Figure .
Figure 9

Illustration of a setup for DR measurement from a semi-infinite tissue with a PD.

Illustration of a setup for DR measurement from a semi-infinite tissue with a PD. By measuring two wavelengths, we found a common point at a specific distance from the light source, as presented in Figure a. The common light intensity for two different wavelengths clearly represents the area that does not depend on the scattering properties.
Figure 10

(a) DR measurements of phantom using two wavelengths; a crossing point appears at 5.1 mm (crosses for 650 nm and squares for 780 nm). (b) In vivo DR measurements of mouse (in inset) using two wavelengths; a crossing point appears at 3.4 mm (crosses for 650 nm and squares for 780 nm).

(a) DR measurements of phantom using two wavelengths; a crossing point appears at 5.1 mm (crosses for 650 nm and squares for 780 nm). (b) In vivo DR measurements of mouse (in inset) using two wavelengths; a crossing point appears at 3.4 mm (crosses for 650 nm and squares for 780 nm). In addition, we measured a semi-infinite medium in vivo. The light intensity as a function of distance from the light source, which was reflected from the carotid artery of mice, is presented in Figure b. The common light intensity point for the mice appears closer to the light source than that in the control phantom. This is expected because the absorption through tissues is higher in the red range than that in the NIR range.[20] Hence, the phantom measurements reveal the IPL point where scattering changes have a minimal impact without absorption components. By detecting this location, we can achieve accurate extraction of absorption coefficients from in vivo measurements, whereas the measurement at the right point excludes the scattering.

Discussion

We claim that the finding of an IPL point in cylindrical (Figure a) and semi-infinite (Figure b) tissues is actually the same phenomenon. Both cases are geometric positions that depend on the distance from the light source where the optical path length is not affected by the optical parameters.
Figure 11

Schemas of (a) an irradiated cylindrical tissue cross section and the location of θIPL, (b) an irradiated semi-infinite tissue and the location of rIPL, (c) both geometries on the same coordinate system, (d) comparison between theoretical and experimental results; the relation between θIPL and tissue radius a is calculated by intersection with rIPL (curve) and experimental results of human fingers (dots).

Schemas of (a) an irradiated cylindrical tissue cross section and the location of θIPL, (b) an irradiated semi-infinite tissue and the location of rIPL, (c) both geometries on the same coordinate system, (d) comparison between theoretical and experimental results; the relation between θIPL and tissue radius a is calculated by intersection with rIPL (curve) and experimental results of human fingers (dots). The relation between the angle of an IPL point in a cylindrical medium, θIPL, and the radius a will be obtained merely by geometrical computations. Let us consider both geometries on the same coordinate system (Figure c). In the semi-infinite geometry, the IPL distance, rIPL, from the illumination point inside the medium defines the following circle equation (blue-dashed line in Figure b)In the cylindrical geometry, the tissue cross section is placed adjacent to the light source (black circle in Figure a); hence, the surface is defined byWe calculated the intersection between eqs and 8 and expressed it in terms of the IPL angle, θIPL, using the Taylor approximation for arccosineIn Figure d, the black line represents this equation for rIPL = 6.4 mm and the relevant range of a. This relation is compared with the experimental results of human fingers where θIPL was measured by a single wavelength by comparing two perpendicular orientations (dots in Figure d from ref (22)). The relation from eq shows a high fit to these experimental results. Hence, the intersection found in eq reveals the θIPL angle in which the IPL point is located according to the finger diameter. Note that the rIPL that presented high correlation to the in vivo results from ref (22) is a value similar to rIPL = 6.8 mm from diffusion theory. Hence, we can determine that the IPL points, which are not influenced by the scattering properties of the medium in the two mentioned geometries, are converged to the same distance from the light source (Supporting Information, Video file).

Conclusions

The optical measurements of tissues are influenced by the absorption as well as scattering. Therefore, there is a challenge to extract an accurate physiological parameter from the absorption regardless of the additional optical parameters’ effect. By using the FSP method, we show the effect of optical properties on the FSP and the IPL point that is not influenced by the reduced scattering coefficient in a cylindrical medium. We present in this article the existence of the IPL phenomenon in theory, simulations, and experimental results using tissuelike phantoms and finally in vivo measurements. In oxygen tissue assessment, the transmitted light is described by the modified Beer–Lambert law,[21]I = I0 e–(μ, where l is the optical path length that depends on the absorption and scattering coefficients and G is a geometric factor. When measuring at several wavelengths, l changes. The existing methods suffer from an inherent error due to these differences in the optical path length, which must be calibrated externally via in vitro measurements. The calibration process adds an error that is too high for some clinical problems;[23,24] therefore, it is challenging to extract the needed parameter (e.g., SpO2) correctly.[25] The IPL point overcomes this problem because the different sources undergo the same optical path length even though they experience different reduced scattering coefficients. Moreover, this phenomenon of a geometric location, where the intensity is independent of the reduced scattering coefficients, appears in a semi-infinite medium as well. We demonstrated that this phenomenon occurs through diffusion theory, MC simulation, and experiments using phantoms as well as in vivo measurements in the DR system. There are researchers who examined the light scattering on the basis of diffusion theory and an equivalent isotropic source and an extrapolated boundary,[17] and their solution agrees with ours. In addition, a similar geometric phenomenon was reported by others[13] about a distance from the light source that is neutralized by the effect of the phase function. This optical parameter has an influence on calculations of the reflected intensity beside a specific location. This location can also be called an IPL point where reflectance is not dependent on the phase function. The correct choice of a measurement point allows the neutralization of the scattering effect and as a result enables us to measure other optical parameters accurately. The use of this unique behavior can predict more accurate measurements in the biomedical field. The measurement of the light intensity at the IPL point, which is independent of the tissue scattering, can improve accuracy in two ways: as a self-calibration point for another measurement point on the FSP in a single wavelength or by measuring the absorption at the IPL point that neutralizes the scattering at two wavelengths.

Computational Methods and Experimental Section

Monte Carlo Simulation

The most common model of simulating light–tissue interactions is the MC simulation. MC simulations are recommended for regions close to the source. We investigated the dependency of the IPL point in the case of different tissue diameters. As we explained in our previous work,[26] we simulated a beam of photons that enters the circular cross section of tissue parallel to the z direction. The propagation path of each photon is calculated from the scattering constant, assuming that the absorption is negligible. The values of the reduced scattering coefficients (μs′) were in the range of the human skin values.[6] In addition, we used a MC simulation from a semi-infinite medium that Prof. Jacques shared with us to deepen and enrich the research. We simulated the FSP of a circular cross section of a tissue via MC method with relevant assumptions and changes. We chose several reduced scattering coefficients and different tissue diameters. This simulation is based on the assumption that all photons reaching the tissue begin as ballistic photons. At each step (dr) of the photon, the new direction (θnew) was calculated using the current photon’s direction (θold) and the probability a photon would scatter, 1 – exp(−μs′dr). In the case of a scattering event, its new direction was calculated using θnew = θold + s × cos–1 g, where s is a random number from the group [−1, 1]. The location of a photon that arrived to the boundary of the cylindrical tissue was saved as the exit angle.

Experiments

FSP Setup

The setup includes a continuous wave illumination using a He–Ne gas laser (Thorlabs, Newton, SA) with a wavelength of λ = 632.8 nm, a maximum power of 8 mW, and a laser beam waist of ω0 = 0.63 mm. In addition, for in vivo measurement, a pigtailed NIR laser (Oxxius S.A., France) was used with λ = 785 nm. We use a portable fixed gain silicon detector as a PD. The PD has an active area of 0.8 mm2, and it is placed on a rotation stage to enable the full scattering profile measurements. The stage can be rotated around a tube with an accuracy of 1°. The voltage is measured every 5°, where 0° is the full transmission and 180° is the full reflection.

Phantoms

Cylindrical tissuelike phantoms were prepared on the basis of intra-lipid (IL) in various concentrations to measure the influence of the reduced scattering coefficient on the FSP experimentally. The IL concentration is calculated according to the reduced scattering coefficients for each phantom,[13] in the range of the human skin values of 8–26 cm–1, similar to that in the MC simulation. The liquid phantoms were placed in test tubes with different diameters. They were measured from 0 to 145°, whereas the remaining angles were not measurable because of the blocking of the light source by the detector.

DR Measurements System

The diffusion reflectance (DR) method is a simple and safe technology for extracting optical parameters using the light intensity at different distances ρ (light source–detector separation). The DR system (NEGOH-OP Technologies, Israel), as was previously described by us,[27] contains two sources, 650 and 780 nm using an optic fiber with a diameter of 125 μm. A portable photodiode was deposited at different distances (ρ) from the source on the sample surface in a micrometer scale. The photodiode had a cross-section diameter of 1 mm2, the reflected intensity was measured. The in vivo measurements using the DR system were carried out on a C57BL male mouse (aged 4.5 months). The mouse was anesthetized by injecting intraperitoneally 200 μL of a mixture of ketamine (10 mg/mL) and xylazine (2 mg/mL). The hair on the neck facing the laser was removed before the measurements. This study received institutional approval from the Bar Ilan Institute for Nanotechnology and Advanced Materials Institutional Animal Care and Use Committee. All in vivo measurements were performed under appropriate anesthesia: the mice barrier-controlled facility followed the guidelines of Bar Ilan’s Institute for Nanotechnology and Advanced Materials Animal Care and Use Committee. The mice were inspected daily by the veterinarian, who handled the appropriate tests and treatment protocols as required. All research protocols were followed strictly by the veterinarian. All major procedures were performed in the surgical facilities using general anesthesia and standard aseptic surgical techniques.
  24 in total

Review 1.  Non-invasive in vivo characterization of breast tumors using photon migration spectroscopy.

Authors:  B J Tromberg; N Shah; R Lanning; A Cerussi; J Espinoza; T Pham; L Svaasand; J Butler
Journal:  Neoplasia       Date:  2000 Jan-Apr       Impact factor: 5.715

2.  Optical properties of selected native and coagulated human brain tissues in vitro in the visible and near infrared spectral range.

Authors:  A N Yaroslavsky; P C Schulze; I V Yaroslavsky; R Schober; F Ulrich; H J Schwarzmaier
Journal:  Phys Med Biol       Date:  2002-06-21       Impact factor: 3.609

3.  Double-integrating-sphere system for measuring the optical properties of tissue.

Authors:  J W Pickering; S A Prahl; N van Wieringen; J F Beek; H J Sterenborg; M J van Gemert
Journal:  Appl Opt       Date:  1993-02-01       Impact factor: 1.980

4.  Reflected light intensity profile of two-layer tissues: phantom experiments.

Authors:  Rinat Ankri; Haim Taitelbaum; Dror Fixler
Journal:  J Biomed Opt       Date:  2011-08       Impact factor: 3.170

Review 5.  Optical properties of biological tissues: a review.

Authors:  Steven L Jacques
Journal:  Phys Med Biol       Date:  2013-05-10       Impact factor: 3.609

Review 6.  Real-time near-infrared fluorescence guided surgery in gynecologic oncology: a review of the current state of the art.

Authors:  Henricus J M Handgraaf; Floris P R Verbeek; Quirijn R J G Tummers; Leonora S F Boogerd; Cornelis J H van de Velde; Alexander L Vahrmeijer; Katja N Gaarenstroom
Journal:  Gynecol Oncol       Date:  2014-08-11       Impact factor: 5.482

Review 7.  Influence of the phase function in generalized diffuse reflectance models: review of current formalisms and novel observations.

Authors:  Katherine W Calabro; Irving J Bigio
Journal:  J Biomed Opt       Date:  2014       Impact factor: 3.170

8.  Simulation of oxygen saturation measurement in a single blood vein.

Authors:  Hamootal Duadi; Meir Nitzan; Dror Fixler
Journal:  Opt Lett       Date:  2016-09-15       Impact factor: 3.776

9.  Estimation of optical pathlength through tissue from direct time of flight measurement.

Authors:  D T Delpy; M Cope; P van der Zee; S Arridge; S Wray; J Wyatt
Journal:  Phys Med Biol       Date:  1988-12       Impact factor: 3.609

10.  Near-infrared human finger measurements based on self-calibration point: Simulation and in vivo experiments.

Authors:  Hamootal Duadi; Idit Feder; Dror Fixler
Journal:  J Biophotonics       Date:  2017-12-21       Impact factor: 3.207

View more
  1 in total

1.  Source-Detector Spectral Pairing-Related Inaccuracies in Pulse Oximetry: Evaluation of the Wavelength Shift.

Authors:  Olivier Tsiakaka; Benoit Gosselin; Sylvain Feruglio
Journal:  Sensors (Basel)       Date:  2020-06-10       Impact factor: 3.576

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.