How the S-ICD (subcutaneous implantable cardiac defibrillator) senses cardiac signals to minimize cardiac over-sensing and maximize rhythm discrimination.

Amy Brisben1.   

Abstract

Transvenous Implantable Cardioverter Defibrillator (TV-ICD) and subcutaneous ICD (S-ICD) devices are available for patients at risk of sudden cardiac death (SCD). Both devices sense ventricular arrhythmias with near 100% sensitivity (Gold et al., 2012) and have similar rates of inappropriate shocks (IAS) (2.5-4.8%). IAS due to supraventricular tachycardias (SVT) is higher for TV-ICDs and IAS due to cardiac over-sensing is higher for S-ICDs (Basu-Ray et al., 2017). The S-ICD's sensing signal is morphologically rich, resembling the surface electrocardiogram (ECG) signal. As such, correlation waveform analysis (CWA) is used by the S-ICD to compare the morphology of two signals to minimize S-ICD cardiac over-sensing and rhythm discrimination errors. Early experience showed that the S-ICD had a comparable overall incidence of IAS compared to TV-ICDs. Progressive improvements in S-ICD algorithms and programming strategies have resulted in overall IAS rate reduction to 4.3% (Theuns et al., 2018), substantiating the S-ICD as a viable choice for patients at risk of SCD.
Copyright © 2018 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2018        PMID: 30220546     DOI: 10.1016/j.jelectrocard.2018.08.022

Source DB:  PubMed          Journal:  J Electrocardiol        ISSN: 0022-0736            Impact factor:   1.438


  4 in total

1.  Twiddler's syndrome with a subcutaneous implantable cardioverter-defibrillator presenting with an inappropriate shock: a case report.

Authors:  C Fielder Camm; Kim Rajappan; Mark Curson; Lindsey Tilling
Journal:  Eur Heart J Case Rep       Date:  2019-11-07

2.  Two Limitations of Subcutaneous Implantable Cardioverter Defibrillator in the Same Patient Warranting Its Explant.

Authors:  Rahul Dhawan; Mansoor Ahmad; Aravdeep Jhand; Sumera Kanwal; Adeel Jamil; Faris Khan
Journal:  Am J Case Rep       Date:  2021-04-29

3.  Ventricular tachycardia detection with the subcutaneous implantable cardioverter-defibrillator: Tracing the cause of an absent tracing.

Authors:  Fadi Mansour; Paul Khairy
Journal:  HeartRhythm Case Rep       Date:  2019-11-02

Review 4.  Successful Identification of and Discrimination Between Atrial and Ventricular Arrhythmia with the Aid of Pacing and Defibrillator Devices.

Authors:  Rahul K Mukherjee; Manav Sohal; Nesan Shanmugam; Simon Pearse; Fadi Jouhra
Journal:  Arrhythm Electrophysiol Rev       Date:  2021-12
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.