| Literature DB >> 30211254 |
Jae Heung Bae1, Sang Yub Lee1,2.
Abstract
PURPOSE: To assess the filter tilting and outcomes of the Celect and Denali inferior vena cava (IVC) filters by using a propensity score-matching analysis.Entities:
Keywords: 3D, three-dimensional; CT, computed tomography; Denali filter; Filter tilt; IVC, inferior vena cava; Inferior vena cava filter; PACS, picture archiving and communication system; Propensity score; Retrieval rate
Year: 2018 PMID: 30211254 PMCID: PMC6134324 DOI: 10.1016/j.ejro.2018.09.001
Source DB: PubMed Journal: Eur J Radiol Open ISSN: 2352-0477
Fig. 1Flow chart shows patient enrollment of this study.
Before and after propensity score adjustment.
| Before Propensity Score Adjustment | After Propensity Score Adjustment | |||||
|---|---|---|---|---|---|---|
| Celect (n = 141) | Denali (n = 43) | Celect (n = 31) | Denali (n = 31) | |||
| Age (years ± mean) | 63.1 ± 14.9 | 60.7 ± 15 | 0.360 | 64.7 ± 11.9 | 63.0 ± 16.0 | 0.481 |
| Sex (male:female) | 73:68 | 27:16 | 0.205 | 20:11 | 17:14 | 0.345 |
| IVC long diameter (mm) | 22.9 ± 3.6 | 23.4 ± 3.1 | 0.425 | 23.5 ± 4.0 | 23.3 ± 3.3 | 0.855 |
| IVC short diameter (mm) | 16.0 ± 3.7 | 16.8 ± 3.1 | 0.210 | 16.2 ± 2.6 | 16.5 ± 3.2 | 0.745 |
| Filter tilt (degrees) | 8.3 ± 5.4 | 5.3 ± 5.9 | 9.5 ± 5.9 | 5.6 ± 6.7 | ||
| Filter tip abutment or penetration | 52/141(37%) | 5/43(12%) | 12/31(39%) | 4/31(13%) | ||
| Abutment:penetration | 25:7 | 4:1 | 12:0 | 3:1 | ||
| Vertical position (cm) | 5.9 ± 1.2 | 5.5 ± 1.7 | 5.7 ± 1.1 | 5.7 ± 0.8 | 0.851 | |
| IVC angulation (degrees) | 9.8 ± 7.4 | 7.7 ± 6.8 | 0.991 | 8.8 ± | 8.8 ± 6.6 | 0.253 |
| 0–10° | 95 | 23 | 21 | 21 | ||
| <10–20° | 32 | 17 | 7 | 7 | ||
| <20° | 14 | 3 | 3 | 3 | ||
| Access site (IJ/RF/LF) | 136/5/0 | 6/21/16 | NA | 30/1/0 | 3/15/13 | NA |
| Retrieval attempt | 100/141 (71%) | 41/43 (93%) | 22/31(71%) | 29/31(94%) | ||
| Retrieval success | 96/100 (96%) | 40/41 (98%) | 1.000 | 22/31(71%) | 29/29(100%) | |
| Retrieval failure | 4/100 (4%) | 1/41 (2%) | 1.000 | 0/22 (0%) | 0/29 (0%) | |
| Advanced or complicated retrieval | 5/100 (5%) | 1/41 (2%) | 0.672 | 0/22 (0%) | 1/29 (3%) | |
| Indwelling time (days, mean ± SD) | 27 ± 24 | 26 ± 21 | 0.893 | 26 ± 22 | 27 ± 23 | 0.746 |
IVC, inferior vena cava; NA, not applicable; IJ, right internal jugular venous access; RF, right common femoral venous access; LF, left common femoral venous access; SD, standard deviation.
Significance of bold values = p value less than 0.05.
Independent t-test.
Chi-square test.
Paired t-test.
McNemar’s test.
Filter retrieval failure and advanced retrieval/complicated retrievals.
| Filter | Age | Sex | IVC diameter (mm) | IVC angle (degrees) | Proximity to the renal vein (cm) | Filter tilt (degrees) | Tip abutment or penetration | Indwelling time (days) | Remarks | |
|---|---|---|---|---|---|---|---|---|---|---|
| Retrieval failure | Celect | 46 | M | 27.7 | 4.8 | 6.0 | 12.1 | Penetration | 8 | Embedded filter tip |
| Retrieval failure | Celect | 33 | M | 21.5 | 3.2 | 1.8 | 1.3 | None | 28 | Thrombotic occlusion of IVC |
| Retrieval failure | Celect | 82 | F | 20.1 | 26.5 | 5.8 | 15.2 | Penetration | 36 | Embedded filter tip |
| Retrieval failure | Celect | 71 | M | 18.5 | 2.8 | 7.5 | 9.3 | Abutment | 13 | Entrapped thrombus |
| Retrieval failure | Denali | 50 | M | 24.8 | 12.9 | 6.3 | 10.2 | None | 10 | Entrapped thrombus |
| Advanced retrieval | Celect | 24 | M | 24.3 | 2.8 | 4.4 | 20.9 | Abutment | 27 | Urokinase + balloon assisted |
| Advanced retrieval | Celect | 64 | M | 27.0 | 7.2 | 5.7 | 13.9 | Penetration | 14 | Balloon assisted |
| Advanced retrieval | Denali | 23 | M | 27.6 | 3.1 | 7.5 | 38.7 | Penetration | 34 | Balloon and forceps assisted |
| Complicated retrieval | Celect | 64 | M | 26.9 | 5.6 | 6.1 | 16.5 | Abutment | 14 | Concealed IVC rupture |
| Complicated retrieval | Celect | 67 | M | 23.2 | 3.4 | 5.5 | 2.1 | None | 15 | Concealed IVC rupture |
| Complicated retrieval | Celect | 71 | F | 21.3 | 7.2 | 6.4 | 10.9 | Abutment | 21 | Concealed IVC rupture |
IVC, inferior vena cava.
Fig. 2Twenty-four-year–old male patient shows IVC filter tilting. A, B. Coronal reformatted computed tomography and venography show IVC filter abutment to the IVC wall. C. A 12-mm balloon catheter was used for reposition of the tilted filter tip. D. Finally, IVC filter was successfully captured with endovascular snare.