Floris E A Udink Ten Cate1,2, Rodina Sobhy3, Atul Kalantre4, Sakshi Sachdev5, Anand Subramanian6, Nageswara Rao Koneti5, Raman Krishna Kumar4, Hala Hamza3, Mahimarangaiah Jayranganath6, Narayanswami Sreeram1. 1. Department of Pediatric Cardiology, Heart Center Cologne, University Hospital of Cologne, Cologne, Germany. 2. Department of Pediatric Cardiology, Amalia Children's Hospital, Radboud University Medical Center, Nijmegen, The Netherlands. 3. Department of Pediatric Cardiology, Cairo University Children's Hospital, Cairo, Egypt. 4. Department of Pediatric Cardiology, Amrita Institute of Medical Sciences, Kochi, India. 5. Department of Pediatric Cardiology, CARE Hospital, Hyderabad, India. 6. Sri Jayadeva Institute of Cardiovascular Sciences and Research, Bengaluru, India.
Abstract
OBJECTIVE: We sought to evaluate the feasibility, technical aspects, and outcome of transcatheter perimembranous ventricular septal defect (pmVSD) closure using duct occluder devices with a single retention disc. BACKGROUND: Use of duct occluder devices to close pmVSD seems a promising alternative therapy. However, limited data exist on this technique. METHODS: From 2010 to 2016, 222 patients (female 47.7%) were identified from databases of five participating institutions in whom pmVSD closure was attempted using an Amplatzer Duct Occluder I or Lifetech duct occluder device. RESULTS: Patients ranged in age from 0.7 to 52 years (median, 7.0 years) and in weight from 4.0 to 70 kg (median, 18.0 kg). The mean size of the VSD was 6.8 ± 2.2 mm. A large defect (> 6 mm) was present in 137 patients (61.7%). Device closure was successful in 218 patients (98.2%). The 10/8 mm device was used in most patients (n = 85, 38.3%), and the vascular approach was from the femoral vein in 169 patients (76.1%). There were 18 early complications in 17/218 patients (7.8%). Three patients (1.4%) developed complete heart block (transient n = 2; requiring permanent pacing n = 1). Median follow-up was 6 months (6 months-6 years). A mild residual shunt was seen in 10 patients at 6 months follow-up. CONCLUSIONS: The immediate results of transcatheter pmVSD closure using a duct occluder device with a single retention disc are promising. It is an effective technique with a lower rate of complications than for other currently available devices.
OBJECTIVE: We sought to evaluate the feasibility, technical aspects, and outcome of transcatheter perimembranous ventricular septal defect (pmVSD) closure using duct occluder devices with a single retention disc. BACKGROUND: Use of duct occluder devices to close pmVSD seems a promising alternative therapy. However, limited data exist on this technique. METHODS: From 2010 to 2016, 222 patients (female 47.7%) were identified from databases of five participating institutions in whom pmVSD closure was attempted using an Amplatzer Duct Occluder I or Lifetech duct occluder device. RESULTS:Patients ranged in age from 0.7 to 52 years (median, 7.0 years) and in weight from 4.0 to 70 kg (median, 18.0 kg). The mean size of the VSD was 6.8 ± 2.2 mm. A large defect (> 6 mm) was present in 137 patients (61.7%). Device closure was successful in 218 patients (98.2%). The 10/8 mm device was used in most patients (n = 85, 38.3%), and the vascular approach was from the femoral vein in 169 patients (76.1%). There were 18 early complications in 17/218 patients (7.8%). Three patients (1.4%) developed complete heart block (transient n = 2; requiring permanent pacing n = 1). Median follow-up was 6 months (6 months-6 years). A mild residual shunt was seen in 10 patients at 6 months follow-up. CONCLUSIONS: The immediate results of transcatheter pmVSD closure using a duct occluder device with a single retention disc are promising. It is an effective technique with a lower rate of complications than for other currently available devices.