A Monte-Carlo-based and GPU-accelerated 4D-dose calculator for a pencil beam scanning proton therapy system.

Mark D Pepin1, Erik Tryggestad1, Hok Seum Wan Chan Tseung1, Jedediah E Johnson1, Michael G Herman1, Chris Beltran1.   

Abstract

PURPOSE: The presence of respiratory motion during radiation treatment leads to degradation of the expected dose distribution, both for target coverage and healthy tissue sparing, particularly for techniques like pencil beam scanning proton therapy which have dynamic delivery systems. While tools exist to estimate this degraded four-dimensional (4D) dose, they typically have one or more deficiencies such as not including the particular effects from a dynamic delivery, using analytical dose calculations, and/or using nonphysical dose-accumulation methods. This work presents a clinically useful 4D-dose calculator that addresses each of these shortcomings.
METHODS: To quickly compute the 4D dose, the three main tasks of the calculator were run on graphics processing units (GPUs). These tasks were (a) simulating the delivery of the plan using measured delivery parameters to distribute the plan amongst 4DCT phases characterizing the patient breathing, (b) using an in-house Monte Carlo simulation (MC) dose calculator to determine the dose delivered to each breathing phase, and (c) accumulating the doses from the various breathing phases onto a single phase for evaluation. The accumulation was performed by individually transferring the energy and mass of dose-grid subvoxels, a technique that models the transfer of dose in a more physically realistic manner. The calculator was run on three test cases, with lung, esophagus, and liver targets, respectively, to assess the various uncertainties in the beam delivery simulation as well as to characterize the dose-accumulation technique.
RESULTS: Four-dimensional doses were successfully computed for the three test cases with computation times ranging from 4-6 min on a server with eight NVIDIA Titan X graphics cards; the most time-consuming component was the MC dose engine. The subvoxel-based dose-accumulation technique produced stable 4D-dose distributions at subvoxel scales of 0.5-1.0 mm without impairing the total computation time. The uncertainties in the beam delivery simulation led to moderate variations of the dose-volume histograms for these cases; the variations were reduced by implementing repainting or phase-gating motion mitigation techniques in the calculator.
CONCLUSIONS: A MC-based and GPU-accelerated 4D-dose calculator was developed to estimate the effects of respiratory motion on pencil beam scanning proton therapy treatments. After future validation, the calculator could be used to assess treatment plans and its quick runtime would make it easily usable in a future 4D-robust optimization system.
© 2018 American Association of Physicists in Medicine.

Entities:  

Keywords:  zzm321990GPUzzm321990; 4D dose calculation; Monte Carlo simulation; proton therapy

Mesh:

Year:  2018        PMID: 30203550     DOI: 10.1002/mp.13182

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  7 in total

1.  Intensity-modulated proton therapy (IMPT) interplay effect evaluation of asymmetric breathing with simultaneous uncertainty considerations in patients with non-small cell lung cancer.

Authors:  Jie Shan; Yunze Yang; Steven E Schild; Thomas B Daniels; William W Wong; Mirek Fatyga; Martin Bues; Terence T Sio; Wei Liu
Journal:  Med Phys       Date:  2020-10-13       Impact factor: 4.071

Review 2.  Proton Therapy for Prostate Cancer: Challenges and Opportunities.

Authors:  Darren M C Poon; Stephen Wu; Leon Ho; Kin Yin Cheung; Ben Yu
Journal:  Cancers (Basel)       Date:  2022-02-13       Impact factor: 6.639

Review 3.  Management of Motion and Anatomical Variations in Charged Particle Therapy: Past, Present, and Into the Future.

Authors:  Julia M Pakela; Antje Knopf; Lei Dong; Antoni Rucinski; Wei Zou
Journal:  Front Oncol       Date:  2022-03-09       Impact factor: 6.244

4.  Functional avoidance-based intensity modulated proton therapy with 4DCT derived ventilation imaging for lung cancer.

Authors:  Jingjing M Dougherty; Edward Castillo; Richard Castillo; Austin M Faught; Mark Pepin; Sean S Park; Chris J Beltran; Thomas Guerrero; Inga Grills; Yevgeniy Vinogradskiy
Journal:  J Appl Clin Med Phys       Date:  2021-06-22       Impact factor: 2.102

5.  Technical Note: Multiple energy extraction techniques for synchrotron-based proton delivery systems may exacerbate motion interplay effects in lung cancer treatments.

Authors:  James E Younkin; Danairis Hernandez Morales; Jiajian Shen; Xiaoning Ding; Joshua B Stoker; Nathan Y Yu; Terence T Sio; Thomas B Daniels; Martin Bues; Mirek Fatyga; Steven E Schild; Wei Liu
Journal:  Med Phys       Date:  2021-07-29       Impact factor: 4.506

Review 6.  Roadmap: proton therapy physics and biology.

Authors:  Harald Paganetti; Chris Beltran; Stefan Both; Lei Dong; Jacob Flanz; Keith Furutani; Clemens Grassberger; David R Grosshans; Antje-Christin Knopf; Johannes A Langendijk; Hakan Nystrom; Katia Parodi; Bas W Raaymakers; Christian Richter; Gabriel O Sawakuchi; Marco Schippers; Simona F Shaitelman; B K Kevin Teo; Jan Unkelbach; Patrick Wohlfahrt; Tony Lomax
Journal:  Phys Med Biol       Date:  2021-02-26       Impact factor: 4.174

7.  AAPM Task Group Report 290: Respiratory motion management for particle therapy.

Authors:  Heng Li; Lei Dong; Christoph Bert; Joe Chang; Stella Flampouri; Kyung-Wook Jee; Liyong Lin; Michael Moyers; Shinichiro Mori; Joerg Rottmann; Erik Tryggestad; Sastry Vedam
Journal:  Med Phys       Date:  2022-01-31       Impact factor: 4.506

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.